Celii, Francis G. and Whitmore, Paul M. and Janda, Kenneth C. (1986) Bakeout controller for the use of helium closed-cycle refrigerators in ultrahigh vacuum applications. Journal of Vacuum Science and Technology A, 4 (4). pp. 1939-1941. ISSN 0734-2101 http://resolver.caltech.edu/CaltechAUTHORS:CELjvsta86
|
PDF
See Usage Policy. 178Kb |
Use this Persistent URL to link to this item: http://resolver.caltech.edu/CaltechAUTHORS:CELjvsta86
Abstract
imple control electronics are described which provide automated bakeout of an ultrahigh vacuum (UHV) chamber equipped with a helium closed-cycle refrigerator. By preventing the refrigerator from reaching elevated temperatures (T>50 °C) during bakeout, these devices can be utilized in the ultrahigh vacuum environment.
| Item Type: | Article |
|---|---|
| Additional Information: | © 1986 American Vacuum Society. We are grateful to Professor Sten Samson for advice on low-temperature techniques, as well as for the thermocouple wire and the loan of a Si diode system for temperature calibration. This work was supported by the United States Office of Naval Research. Arthur Amos Noyes Laboratory of Chemical Physics, Contribution No. 7236. [F.G.C. was an] Atlantic Richfield Foundation Fellow. [K.C.J. was a] Camille and Henry Dreyfus Foundation Fellow. |
| Record Number: | CaltechAUTHORS:CELjvsta86 |
| Persistent URL: | http://resolver.caltech.edu/CaltechAUTHORS:CELjvsta86 |
| Alternative URL: | http://dx.doi.org/10.1116/1.573750 |