Variable Sensitivity Automatic Ionization Chamber

Neher, H. V. (1961) Variable Sensitivity Automatic Ionization Chamber. Review of Scientific Instruments, 32 (1). pp. 48-49. ISSN 0034-6748. doi:10.1063/1.1717143. https://resolver.caltech.edu/CaltechAUTHORS:NEHrsi61

[img]
Preview
PDF - Published Version
See Usage Policy.

116kB

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:NEHrsi61

Abstract

An all quartz electroscope system is described whose sensitivity may be varied over a wide range by changing one potential. Like other previously described systems, it accumulates a definite quantity of electric charge, for given potentials, and then automatically recharges itself. When this recharging occurs an electrical pulse may be taken off for recording purposes. It is possible to change the sensitivity by a factor of 500.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1063/1.1717143DOIUNSPECIFIED
http://link.aip.org/link/?RSINAK/32/48/1PublisherUNSPECIFIED
Additional Information:©1961 The American Institute of Physics. Received November 7, 1960. This development was made in connection wit research work sponsored jointly by the Office of Naval Research and the U.S. Atomic Energy Commission. This assistance is gratefully recognized.
Funders:
Funding AgencyGrant Number
Office of Naval ResearchUNSPECIFIED
Atomic Energy CommissionUNSPECIFIED
Issue or Number:1
DOI:10.1063/1.1717143
Record Number:CaltechAUTHORS:NEHrsi61
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:NEHrsi61
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.