CaltechAUTHORS
  A Caltech Library Service

A study of the autofluorescence of parylene materials for µTAS applications

Lu, Bo and Zheng, Siyang and Quach, Brandon Quoc and Tai, Yu-Chong (2010) A study of the autofluorescence of parylene materials for µTAS applications. Lab on a Chip, 10 . pp. 1826-1834. ISSN 1473-0197. https://resolver.caltech.edu/CaltechAUTHORS:20100521-110030265

[img]
Preview
PDF - Published Version
See Usage Policy.

461Kb

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20100521-110030265

Abstract

Parylene-C has been widely used as a biocompatible material for microfluidics and micro total analysis system (µTAS) applications in recent decades. However, its autofluorescence can be a major obstacle for parylene-C based devices used in applications requiring sensitive fluorescence detection. In this paper, Parylene-C was compared with other commonly used polymer and plastic materials in µTAS devices for their autofluorescence. We also report here an in-depth study of the behaviors and mechanisms of the autofluorescence of parylene-C, as well as several other commercialized members in the parylene family, including parylene-D, parylene-N and parylene-HT, using epifluorescence microscopy, fluorimeter and infrared spectroscopy. Strong autofluorescence was induced in parylene-C during short-wavelength excitation (i.e. UV excitation). Variation of autofluorescence intensity of parylene-C film was found to be related to both dehydrogenation and photo-oxidation. Moreover, the influence of microfabrication process on parylene-C autofluorescence was also evaluated. Parylene-HT, which exhibits low initial autofluorescence, decreasing autofluorescence behavior under UV excitation and higher UV stability, can be a promising alternative for µTAS applications with fluorescence detection.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1039/b924855bDOIUNSPECIFIED
http://www.rsc.org/publishing/journals/LC/article.asp?doi=b924855bPublisherUNSPECIFIED
ORCID:
AuthorORCID
Tai, Yu-Chong0000-0001-8529-106X
Additional Information:© 2010 The Royal Society of Chemistry. Received 27th November 2009, Accepted 19th March 2010. First published on the web 29th April 2010. The authors would like to thank Beckman Institute Laser Resource Center of Caltech for the spectrofluorimeter, and Dr John Bercaw's group of Caltech for their FT-IR Spectrometer. The author would like to thank Alec Durrell and Alex Miller for equipment training and experiment assistance. Also the authors would like to thank Kuang Shen from the Chemistry Department of Caltech for his helpful discussion and assistance. The funding is provided by NIH, under Award Number 4000072624.
Funders:
Funding AgencyGrant Number
NIH4000072624
Record Number:CaltechAUTHORS:20100521-110030265
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20100521-110030265
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:18387
Collection:CaltechAUTHORS
Deposited By: Tony Diaz
Deposited On:14 Jul 2010 18:21
Last Modified:03 Oct 2019 01:42

Repository Staff Only: item control page