CaltechAUTHORS
  A Caltech Library Service

A low noise, high thermal stability, 0.1 K test facility for the Planck HFI bolometers

Paine, C. G. and Bock, J. J. and Hristov, V. V. and Lange, A. E. (2002) A low noise, high thermal stability, 0.1 K test facility for the Planck HFI bolometers. In: Advances in Cryogenic Engineering. AIP Conference Proceedings. No.613. American Institute of Physics , Melville, N.Y. , pp. 1651-1658. ISBN 0-7354-0059-8. https://resolver.caltech.edu/CaltechAUTHORS:20111109-153507044

[img]
Preview
PDF - Published Version
See Usage Policy.

1549Kb

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20111109-153507044

Abstract

We are developing a facility which will be used to characterize the bolometric detectors for Planck, an ESA mission to investigate the Cosmic Microwave Background. The bolometers operate at 0.1 K, employing neutron-transmutation doped (NTD) Ge thermistors with resistances of several megohms to achieve NEPs~1×10^(–17) W Hz^(–1/2). Characterization of the intrinsic noise of the bolometers at frequencies as low as 0.010 Hz dictates a test apparatus thermal stability of 40 nK Hz^(–1/2) to that frequency. This temperature stability is achieved via a multi-stage isolation and control geometry with high resolution thermometry implemented with NTD Ge thermistors, JFET source followers, and dedicated lock-in amplifiers. The test facility accommodates 24 channels of differential signal readout, for measurement of bolometer V(I) characteristics and intrinsic noise. The test facility also provides for modulated radiation in the submillimeter band incident on the bolometers, for measurement of the optical speed-of-response; this illumination can be reduced below detectable limits without interrupting cryogenic operation. A commercial Oxford Instruments dilution refrigerator provides the cryogenic environment for the test facility.


Item Type:Book Section
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1063/1.1472202DOIUNSPECIFIED
http://link.aip.org/link/?APCPCS/613/1651/1PublisherUNSPECIFIED
ORCID:
AuthorORCID
Bock, J. J.0000-0002-5710-5212
Additional Information:© 2002 American Institute of Physics. Issue Date: 10 May 2002. The research described in this paper was performed at the Jet Propulsion Laboratory, California Institute of Technology, under a contract with the National Aeronautics and Space Administration.
Funders:
Funding AgencyGrant Number
NASAUNSPECIFIED
Subject Keywords:refrigerators, cryogenics, test facilities, bolometers, cosmic background radiation, thermistors
Series Name:AIP Conference Proceedings
Issue or Number:613
Classification Code:PACS: 07.87.+v; 07.20.Mc;
Record Number:CaltechAUTHORS:20111109-153507044
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20111109-153507044
Official Citation:A low noise, high thermal stability, 0.1 K test facility for the Planck HFI bolometers C. G. Paine, J. J. Bock, V. V. Hristov, and A. E. Lange, AIP Conf. Proc. 613, 1651 (2002), DOI:10.1063/1.1472202
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:27717
Collection:CaltechAUTHORS
Deposited By: Ruth Sustaita
Deposited On:09 Nov 2011 23:57
Last Modified:03 Oct 2019 03:25

Repository Staff Only: item control page