Betabox: a beta particle imaging system based on a position sensitive avalanche photodiode

Dooraghi, A. A. and Vu, N. T. and Silverman, R. W. and Farrell, R. and Shah, K. S. and Wang, J. and Heath, J. R. and Chatziioannou, A. F. (2013) Betabox: a beta particle imaging system based on a position sensitive avalanche photodiode. Physics in Medicine and Biology, 58 (11). pp. 3739-3753. ISSN 0031-9155. PMCID PMC3706465. https://resolver.caltech.edu/CaltechAUTHORS:20130621-101448619

[img] PDF - Accepted Version
See Usage Policy.

4Mb

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20130621-101448619

Abstract

A beta camera has been developed that allows planar imaging of the spatial and temporal distribution of beta particles using a 14 × 14 mm^2 position sensitive avalanche photodiode (PSAPD). This camera system, which we call Betabox, can be directly coupled to microfluidic chips designed for cell incubation or other biological applications. Betabox allows for imaging the cellular uptake of molecular imaging probes labeled with charged particle emitters such as ^(18)F inside these chips. In this work, we investigate the quantitative imaging capabilities of Betabox for ^(18)F beta particles, in terms of background rate, efficiency, spatial resolution, and count rate. Measurements of background and spatial resolution are considered both at room temperature (21 °C ± 1 °C) and at an elevated operating temperature (37 °C ± 1 °C), as is often required for biological assays. The background rate measured with a 4 keV energy cutoff is below 2 cph mm^(−2) at both 21 and 37 °C. The absolute efficiency of Betabox for the detection of ^(18)F positron sources in contact with a PSAPD with the surface passivated from ambient light and damage is 46% ± 1%. The lower detection limit is estimated using the Rose Criterion to be 0.2 cps mm^(−2) for 1 min acquisitions and a 62 × 62 µm^2 pixel size. The upper detection limit is approximately 21 000 cps. The spatial resolution at both 21 and 37 °C ranges from 0.4 mm FWHM at the center of the field of view (FOV), and degrades to 1 mm at a distance of 5 mm away from center yielding a useful FOV of approximately 10 × 10 mm^2. We also investigate the effects on spatial resolution and sensitivity that result from the use of a polymer based microfluidic chip. For these studies we place varying layers of low-density polyethylene (LDPE) between the detector and the source and find that the spatial resolution degrades by ~180 µm for every 100 µm of LDPE film. Sensitivity is reduced by half with the inclusion of ~200 µm of additional LDPE film. Lastly, we demonstrate the practical utilization of Betabox, with an imaging test of its linearity, when coupled to a polydimethylsiloxane microfluidic chip designed for cell based assays.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1088/0031-9155/58/11/3739 DOIArticle
http://iopscience.iop.org/0031-9155/58/11/3739/PublisherArticle
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3706465/PubMed CentralArticle
ORCID:
AuthorORCID
Heath, J. R.0000-0001-5356-4385
Additional Information:© 2013 Institute of Physics and Engineering in Medicine. Received 17 December 2012, in final form 6 April 2013. Published 8 May 2013. The authors would like to thank the staff of the Crump Imaging facility at UCLA including Waldemar Ladno, Darin Williams, and Jeff Collins for their assistance with obtaining ^(18)F solution for this study. The authors would also like to thank Hongkai Wang for valuable discussions on image distortion correction and David Prout for assistance on signal processing. Arion Chatziioannou and Nam Vu are coauthors of a patent owned by the University of California that includes the technology described in this work. This work was supported in part by the National Institutes of Health (R25CA098010 and U54CA119347), the Department of Energy (DE-SC0001249) and the UCLA Foundation from a donation made by Ralph and Marjorie Crump for the UCLA Crump Institute for Molecular Imaging.
Funders:
Funding AgencyGrant Number
NIHR25CA098010
NIHU54CA119347
Department of Energy (DOE)DE-SC0001249
UCLA FoundationUNSPECIFIED
UCLA Crump Institute for Molecular ImagingUNSPECIFIED
Issue or Number:11
Classification Code:PACS: 87.85.Lf, 87.63.-d, 85.60.Dw, 87.57.uk
PubMed Central ID:PMC3706465
Record Number:CaltechAUTHORS:20130621-101448619
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20130621-101448619
Official Citation:Betabox: a beta particle imaging system based on a position sensitive avalanche photodiode A A Dooraghi, N T Vu, R W Silverman, R Farrell, K S Shah, J Wang, J R Heath and A F Chatziioannou doi:10.1088/0031-9155/58/11/3739
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:39020
Collection:CaltechAUTHORS
Deposited By: Ruth Sustaita
Deposited On:21 Jun 2013 17:39
Last Modified:03 Oct 2019 05:03

Repository Staff Only: item control page