Kankeleit, E. (1964) Velocity Spectrometer for Mössbauer Experiments. Review of Scientific Instruments, 35 (2). pp. 194-197. ISSN 0034-6748. doi:10.1063/1.1718776. https://resolver.caltech.edu/CaltechAUTHORS:KANrsi64
![]()
|
PDF
See Usage Policy. 415kB |
Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:KANrsi64
Abstract
An electromechanical feedback system for Mössbauer spectroscopy with a velocity range of 2×10^–5 to 60 cm/sec is described. A reference signal of the desired waveform (triangle, parabola, etc.) to drive the feedback system is produced by the multichannel analyzer and with the help of operational amplifiers. The linearity of the system for a triangular velocity waveform is better than 0.5% during 98% of the half-period.
Item Type: | Article | ||||||
---|---|---|---|---|---|---|---|
Related URLs: |
| ||||||
Additional Information: | ©1964 The American Institute of Physics (Received 23 September 1963; revised 23 October 1963) The author wants to thank H. Henrickson for his help in constructing the drive and F. Snively for building the power amplifier. | ||||||
Issue or Number: | 2 | ||||||
DOI: | 10.1063/1.1718776 | ||||||
Record Number: | CaltechAUTHORS:KANrsi64 | ||||||
Persistent URL: | https://resolver.caltech.edu/CaltechAUTHORS:KANrsi64 | ||||||
Usage Policy: | No commercial reproduction, distribution, display or performance rights in this work are provided. | ||||||
ID Code: | 4872 | ||||||
Collection: | CaltechAUTHORS | ||||||
Deposited By: | Archive Administrator | ||||||
Deposited On: | 11 Sep 2006 | ||||||
Last Modified: | 08 Nov 2021 20:20 |