Kim, Chulhong and Erpelding, Todd N. and Jankovic, Ladislav and Wang, Lihong V. (2011) Performance benchmarks of an array-based hand-held photoacoustic probe adapted from a clinical ultrasound system for non-invasive sentinel lymph node imaging. Philosophical Transactions A: Mathematical, Physical and Engineering Sciences, 369 (1955). pp. 4644-4650. ISSN 1364-503X. PMCID PMC3263783. https://resolver.caltech.edu/CaltechAUTHORS:20160708-091728581

Full text is not posted in this repository. Consult Related URLs below.

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20160708-091728581

Abstract

Clinical translation of photoacoustic (PA) imaging can be facilitated by integration with commercial ultrasound (US) scanners to enable dual-modality imaging. An array-based US scanner was modified for hand-held PA imaging. The performance was benchmarked in terms of signal-to-noise ratio (SNR), axial spatial resolution and sensitivity. PA images of a tube, filled with methylene blue (MB; approx. 30 mM) and placed at various depths in chicken tissue, were acquired. A 5 cm penetration depth was achieved with an 18.6 dB SNR using a laser fluence of 3 mJ cm^(−2), only one-seventh of the safety limit (20 mJ cm^(−2)). An axial resolution of approximately 400 μm was maintained at all imaging depths. The PA sensitivity to MB placed 2.3 cm deep in chicken tissue was less than 100 μM. Further, after intradermal injection of MB (approx. 30 mM), a rat sentinel lymph node was clearly identified in vivo, beneath a 3.8 cm thick layer of chicken breast. The accumulated concentration of MB in the node was estimated to be approximately 7 mM. The noise-equivalent sensitivities (approx. 2 cm depth) were 17 and 85 μM, ex vivo and in vivo, respectively. These results support the use of this PA system for non-invasive mapping and image-guided needle biopsy of sentinel nodes in breast cancer patients.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1098/rsta.2010.0353DOIArticle
http://rsta.royalsocietypublishing.org/content/369/1955/4644PublisherArticle
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3263783/PubMed CentralArticle
ORCID:
AuthorORCID
Wang, Lihong V.0000-0001-9783-4383
Additional Information:© 2011 The Royal Society. Published 17 October 2011. This work was supported in part by grants from the National Institutes of Health (R01 EB000712, R01 EB008085, R01 CA134539, U54 CA136398, R01 EB010049 and 5P60 DK02057933). L.V.W. has a financial interest in Microphotoacoustics, Inc. and in Endra, Inc., which, however, did not support this work.
Funders:
Funding AgencyGrant Number
NIHR01 EB000712
NIHR01 EB008085
NIHR01 CA134539
NIHU54 CA136398
NIHR01 EB010049
NIH5P60 DK02057933
Subject Keywords:photoacoustic imaging; clinical ultrasound array system; sentinel lymph node biopsy; image-guided needle biopsy
Issue or Number:1955
PubMed Central ID:PMC3263783
Record Number:CaltechAUTHORS:20160708-091728581
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20160708-091728581
Official Citation:Performance benchmarks of an array-based hand-held photoacoustic probe adapted from a clinical ultrasound system for non-invasive sentinel lymph node imaging Chulhong Kim, Todd N. Erpelding, Ladislav Jankovic, Lihong V. Wang Phil. Trans. R. Soc. A 2011 369 4644-4650; DOI: 10.1098/rsta.2010.0353. Published 17 October 2011
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:68913
Collection:CaltechAUTHORS
Deposited By: Tony Diaz
Deposited On:09 Jul 2016 18:13
Last Modified:03 Oct 2019 10:17

Repository Staff Only: item control page