CaltechAUTHORS
  A Caltech Library Service

Retrospective respiration-gated whole-body photoacoustic computed tomography of mice

Xia, Jun and Chen, Wanyi and Maslov, Konstantin and Anastasio, Mark A. and Wang, Lihong V. (2014) Retrospective respiration-gated whole-body photoacoustic computed tomography of mice. Journal of Biomedical Optics, 19 (1). Art. No. 016003. ISSN 1083-3668. PMCID PMC3881607. https://resolver.caltech.edu/CaltechAUTHORS:20160721-095056076

[img] PDF - Published Version
See Usage Policy.

2111Kb

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20160721-095056076

Abstract

Photoacoustic tomography (PAT) is an emerging technique that has a great potential for preclinical whole-body imaging. To date, most whole-body PAT systems require multiple laser shots to generate one cross-sectional image, yielding a frame rate of <1 Hz. Because a mouse breathes at up to 3 Hz, without proper gating mechanisms, acquired images are susceptible to motion artifacts. Here, we introduce, for the first time to our knowledge, retrospective respiratory gating for whole-body photoacoustic computed tomography. This new method involves simultaneous capturing of the animal’s respiratory waveform during photoacoustic data acquisition. The recorded photoacoustic signals are sorted and clustered according to the respiratory phase, and an image of the animal at each respiratory phase is reconstructed subsequently from the corresponding cluster. The new method was tested in a ring-shaped confocal photoacoustic computed tomography system with a hardware-limited frame rate of 0.625 Hz. After respiratory gating, we observed sharper vascular and anatomical images at different positions of the animal body. The entire breathing cycle can also be visualized at 20 frames/cycle.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1117/1.JBO.19.1.016003DOIArticle
http://biomedicaloptics.spiedigitallibrary.org/article.aspx?articleid=1812778PublisherArticle
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3881607PubMed CentralArticle
ORCID:
AuthorORCID
Wang, Lihong V.0000-0001-9783-4383
Additional Information:© 2014 SPIE. Paper 130724R received Oct. 6, 2013; revised manuscript received Nov. 20, 2013; accepted for publication Nov. 27, 2013; published online Jan. 6, 2014. The authors appreciate Prof. James Ballard’s close reading of the manuscript. This work was sponsored in part by National Institutes of Health grants DP1 EB016986 (NIH Director’s Pioneer Award), R01 EB016963, R01 CA134539, R01 EB010049, and R01 CA159959. L.W. has a financial interest in Microphotoacoustics, Inc. and Endra, Inc., which, however, did not support this work. K.M. has a financial interest in Microphotoacoustics, Inc., which, however, did not support this work.
Funders:
Funding AgencyGrant Number
NIHDP1 EB016986
NIHR01 EB016963
NIHR01 CA134539
NIHR01 EB010049
NIHR01 CA159959
Subject Keywords:photoacoustic computed tomography; small-animal whole-body imaging; respiratory motion; retrospective motion gating
Issue or Number:1
PubMed Central ID:PMC3881607
Record Number:CaltechAUTHORS:20160721-095056076
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20160721-095056076
Official Citation:Xia J, Chen W, Maslov K, Anastasio MA, Wang LV; Retrospective respiration-gated whole-body photoacoustic computed tomography of mice. J. Biomed. Opt. 0001;19(1):016003. doi:10.1117/1.JBO.19.1.016003
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:69147
Collection:CaltechAUTHORS
Deposited By: Tony Diaz
Deposited On:25 Jul 2016 22:20
Last Modified:03 Oct 2019 10:19