Simultaneous fluorescence and high-resolution bright-field imaging with aberration correction over a wide field-of-view with Fourier ptychographic microscopy (FPM)

Chung, Jaebum and Kim, Jinho and Ou, Xiaoze and Horstmeyer, Roarke and Yang, Changhuei (2016) Simultaneous fluorescence and high-resolution bright-field imaging with aberration correction over a wide field-of-view with Fourier ptychographic microscopy (FPM). In: Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XXIII. Proceedings of SPIE. No.9713. Society of Photo-Optical Instrumentation Engineers , Bellingham, WA, Art. No. 97130I. ISBN 978-1-62841-947-4. https://resolver.caltech.edu/CaltechAUTHORS:20160804-095326080

[img] PDF - Published Version
See Usage Policy.

118Kb

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20160804-095326080

Abstract

We present a method to acquire both fluorescence and high-resolution bright-field images with correction for the spatially varying aberrations over a microscope’s wide field-of-view (FOV). First, the procedure applies Fourier ptychographic microscopy (FPM) to retrieve the amplitude and phase of a sample, at a resolution that significantly exceeds the cutoff frequency of the microscope objective lens. At the same time, FPM algorithm is able to leverage on the redundancy within the set of acquired FPM bright-field images to estimate the microscope aberrations, which usually deteriorate in regions further away from the FOV’s center. Second, the procedure acquires a raw wide-FOV fluorescence image within the same setup. Lack of moving parts allows us to use the FPM-estimated aberration map to computationally correct for the aberrations in the fluorescence image through deconvolution. Overlaying the aberration-corrected fluorescence image on top of the high-resolution bright-field image can be done with accurate spatial correspondence. This can provide means to identifying fluorescent regions of interest within the context of the sample’s bright-field information. An experimental demonstration successfully improves the bright-field resolution of fixed, stained and fluorescently tagged HeLa cells by a factor of 4.9, and reduces the error caused by aberrations in a fluorescence image by 31%, over a field of view of 6.2 mm by 9.3 mm. For optimal deconvolution, we show the fluorescence image needs to have a signal-to-noise ratio of ~18.


Item Type:Book Section
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1117/12.2211974DOIArticle
http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2519065PublisherArticle
ORCID:
AuthorORCID
Horstmeyer, Roarke0000-0002-2480-9141
Additional Information:© 2015 SPIE.
Series Name:Proceedings of SPIE
Issue or Number:9713
Record Number:CaltechAUTHORS:20160804-095326080
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20160804-095326080
Official Citation:Jaebum Chung ; Jinho Kim ; Xiaoze Ou ; Roarke Horstmeyer and Changhuei Yang. "Simultaneous fluorescence and high-resolution bright-field imaging with aberration correction over a wide field-of-view with Fourier ptychographic microscopy (FPM) (Conference Presentation)" Proc. SPIE 9713, Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XXIII, 97130I (April 27, 2016). doi:10.1117/12.2211974.
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:69435
Collection:CaltechAUTHORS
Deposited By: Melissa Ray
Deposited On:04 Aug 2016 18:44
Last Modified:03 Oct 2019 10:22

Repository Staff Only: item control page