CaltechAUTHORS
  A Caltech Library Service

Curved array photoacoustic tomographic system for small animal imaging

Gamelin, John and Aguirre, Andres and Maurudis, Anastasios and Huang, Fei and Castillo, Diego and Wang, Lihong V. and Zhu, Quing (2008) Curved array photoacoustic tomographic system for small animal imaging. Journal of Biomedical Optics, 13 (2). Art. No. 024007. ISSN 1083-3668. PMCID PMC2507725. doi:10.1117/1.2907157. https://resolver.caltech.edu/CaltechAUTHORS:20160922-074407860

[img] PDF - Published Version
See Usage Policy.

1MB

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20160922-074407860

Abstract

We present systematic characterization of a photoacoustic imaging system optimized for rapid, high-resolution tomographic imaging of small animals. The system is based on a 128-element ultrasonic transducer array with a 5-MHz center frequency and 80% bandwidth shaped to a quarter circle of 25mm radius. A 16-channel data-acquisition module and dedicated channel detection electronics enable capture of a 90-deg field-of-view image in less than 1s and a complete 360-deg scan using sample rotation within 15s. Measurements on cylindrical phantom targets demonstrate a resolution of better than 200μm and high-sensitivity detection of 580-μm blood tubing to depths greater than 3cm in a turbid medium with reduced scattering coefficient μ′s =7.8cm^(−1). The system is used to systematically investigate the effects of target size, orientation, and geometry on tomographic imaging. As a demonstration of these effects and the system imaging capabilities, we present tomographic photoacoustic images of the brain vasculature of an ex vivo mouse with varying measurement aperture. For the first time, according to our knowledge, resolution of sub-200-μm vessels with an overlying turbid medium of greater than 2cm depth is demonstrated using only intrinsic biological contrast.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1117/1.2907157DOIArticle
http://biomedicaloptics.spiedigitallibrary.org/article.aspx?articleid=1102615PublisherArticle
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2507725PubMed CentralArticle
ORCID:
AuthorORCID
Wang, Lihong V.0000-0001-9783-4383
Additional Information:© 2008 SPIE. Paper 07224R received Jun. 19, 2007; revised manuscript received Sep. 6, 2007; accepted for publication Nov. 6, 2007; published online Apr. 17, 2008. We acknowledge partial support from National Institutes of Health (NIH) grants NIH R01NS46214 and NIH R01EB002136.
Funders:
Funding AgencyGrant Number
NIHR01NS46214
NIHR01EB002136
Subject Keywords:photoacoustics; tomography; biomedical optics; medical imaging; point spread functions; ultrasonics
Issue or Number:2
PubMed Central ID:PMC2507725
DOI:10.1117/1.2907157
Record Number:CaltechAUTHORS:20160922-074407860
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20160922-074407860
Official Citation:Gamelin J, Aguirre A, Maurudis A, et al; Curved array photoacoustic tomographic system for small animal imaging. J. Biomed. Opt. 0001;13(2):024007-024007-10
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:70527
Collection:CaltechAUTHORS
Deposited By: Tony Diaz
Deposited On:30 Sep 2016 18:20
Last Modified:11 Nov 2021 04:30

Repository Staff Only: item control page