CaltechAUTHORS
  A Caltech Library Service

Effects of acoustic heterogeneities on transcranial brain imaging with microwave-induced thermoacoustic tomography

Jin, Xing and Li, Changhui and Wang, Lihong V. (2008) Effects of acoustic heterogeneities on transcranial brain imaging with microwave-induced thermoacoustic tomography. Medical Physics, 35 (7). pp. 3205-3214. ISSN 0094-2405. PMCID PMC2664839. doi:10.1118/1.2938731. https://resolver.caltech.edu/CaltechAUTHORS:20160922-142330642

[img] PDF - Published Version
See Usage Policy.

307kB

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20160922-142330642

Abstract

The effects of acoustic heterogeneities on transcranial brainimaging with microwave-induced thermoacoustic tomography were studied. A numerical model for calculating the propagation of thermoacousticwaves through the skull was developed and experimentally examined. The model takes into account wave reflection and refraction at the skull surfaces and therefore provides improved accuracy for the reconstruction. To evaluate when the skull-induced effects could be ignored in reconstruction, the reconstructed images obtained by the proposed method were further compared with those obtained with the method based on homogeneous acoustic properties. From simulation and experimental results, it was found that when the target region is close to the center of the brain, the effects caused by the skull layer are minimal and both reconstruction methods work well. As the target region becomes closer to the interface between the skull and braintissue, however, the skull-induced distortion becomes increasingly severe, and the reconstructed image would be strongly distorted without correcting those effects. In this case, the proposed numerical method can improve image quality by taking into consideration the wave refraction and mode conversion at the skull surfaces. This work is important for obtaining good brainimages when the thickness of the skull cannot be ignored.


Item Type:Article
Related URLs:
URLURL TypeDescription
http://dx.doi.org/10.1118/1.2938731DOIArticle
http://scitation.aip.org/content/aapm/journal/medphys/35/7/10.1118/1.2938731PublisherArticle
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2664839PubMed CentralArticle
ORCID:
AuthorORCID
Wang, Lihong V.0000-0001-9783-4383
Additional Information:© 2008 American Association of Physicists in Medicine. Received 08 October 2007; Revised 09 May 2008; Accepted 12 May 2008; Published online 19 June 2008. The authors wish to thank the Washington National Primate Research Center for providing the monkey sample, and George Stoica and Alejandra Garcia-Uribe for their advice and help with the tissue sample. We acknowledge the useful comments from the reviewers as well. This project was sponsored by National Institutes of Health Grants R01 NS46214 and R01 EB000712.
Funders:
Funding AgencyGrant Number
NIHR01 NS46214
NIHR01 EB000712
Subject Keywords:thermoacoustic tomography, transcranial brain imaging, acoustic heterogeneities
Issue or Number:7
PubMed Central ID:PMC2664839
DOI:10.1118/1.2938731
Record Number:CaltechAUTHORS:20160922-142330642
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20160922-142330642
Official Citation:Effects of acoustic heterogeneities on transcranial brain imaging with microwave-induced thermoacoustic tomography Jin, Xing and Li, Changhui and Wang, Lihong V., Medical Physics, 35, 3205-3214 (2008), DOI:http://dx.doi.org/10.1118/1.2938731
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.