CaltechAUTHORS
  A Caltech Library Service

Preliminary performance measurements of bolometers for the Planck high-frequency instrument

Holmes, W. and Bock, J. and Ganga, K. and Hristov, V. V. and Hustead, L. and Koch, T. and Lange, A. E. and Paine, C. and Yun, M. (2003) Preliminary performance measurements of bolometers for the Planck high-frequency instrument. In: Millimeter and Submillimeter Detectors for Astronomy. Proceedings of SPIE. No.4855. Society of Photo-optical Instrumentation Engineers (SPIE) , Bellingham, WA, pp. 208-216. ISBN 9780819446343. https://resolver.caltech.edu/CaltechAUTHORS:20181113-081049718

[img] PDF - Published Version
See Usage Policy.

776kB

Use this Persistent URL to link to this item: https://resolver.caltech.edu/CaltechAUTHORS:20181113-081049718

Abstract

We report on the characterization of bolometers fabricated at the Jet Propulsion Laboratory for the High Frequency Instrument (HFI) of the joint ESA/NASA Herschel/Planck mission to be launched in 2007. The HFI is a multicolor focal plane which consists of 48 bolometers operated at 100mK. Each bolometer is mounted to a feedhorn-filter assembly which defines one of six frequency bands centered between 100-857GHz. Four detectors in each of six bands are coupled to both linear polarizations and thus measure the total intensity. In addition, eight detectors in each of 3 bands (143, 217, and 353GHz) couple only to a single linear polarization and thus provide measurements of the Stokes parameters, Q and U, as well the total intensity. The detectors are required to achieve a Noise Equivalent Power (NEP) at or below the background limit ∼ 10^(-17)W/√Hz for the telescope and time constants of a few ms, short enough to resolve point sources as the 5 to 9 arc-minute beams move across the sky in great circles at 1 rpm. The bolometers are tested at 100mK in a commercial dilution refrigerator with a custom built thermal control system to regulate the heat sink with precision < 100nK/√Hz. The 100mK tests include dark electrical characterization of the load curves, optical and electrical measurement of the thermal time constants and measurement of the noise spectral density from 0.01 to 10Hz for up to 24 bolometers simultaneously.


Item Type:Book Section
Related URLs:
URLURL TypeDescription
https://doi.org/10.1117/12.459421DOIArticle
ORCID:
AuthorORCID
Bock, J.0000-0002-5710-5212
Additional Information:© 2003 Society of Photo-optical Instrumentation Engineers (SPIE). This work was performed for the Jet Propulsion Laboratory, California Institute of Technology, sponsored by the National Aeronautics and Space Administration. We acknowledge helpful discussions with Bill Jones.
Funders:
Funding AgencyGrant Number
NASA/JPL/CaltechUNSPECIFIED
Subject Keywords:mm-wave sensors, bolometer, micro-machining, cosmic microwave background
Series Name:Proceedings of SPIE
Issue or Number:4855
DOI:10.1117/12.459421
Record Number:CaltechAUTHORS:20181113-081049718
Persistent URL:https://resolver.caltech.edu/CaltechAUTHORS:20181113-081049718
Official Citation:Warren Holmes, James J. Bock, Ken Ganga, Viktor Hristov, Leonard Hustead, Timothy Koch, Andrew E. Lange, Chris Paine, Minhee Yun, "Preliminary performance measurements of bolometers for the Planck High Frequency Instrument," Proc. SPIE 4855, Millimeter and Submillimeter Detectors for Astronomy, (17 February 2003)
Usage Policy:No commercial reproduction, distribution, display or performance rights in this work are provided.
ID Code:90851
Collection:CaltechAUTHORS
Deposited By: Tony Diaz
Deposited On:13 Nov 2018 17:58
Last Modified:16 Nov 2021 03:36

Repository Staff Only: item control page