Lin et al.
Light: Science & Applications
(2023) 12:12
Of
fi
cial journal of the CIOMP 2047-7538
https://doi.org/10.1038/s41377-022-01053-7
www.nature.com/lsa
ARTICLE
Open Access
Non-invasive photoacoustic computed
tomography of rat heart anatomy and function
Li Lin
1,4,5
, Xin Tong
1
, Susana Cavallero
2
,YideZhang
1
,ShuaiNa
1
,RuiCao
1
,TzungK.Hsiai
2,3
✉
and
Lihong V. Wang
1
✉
Abstract
Complementary to mainstream cardiac imaging modalities for preclinical research, photoacoustic computed
tomography (PACT) can provide functional optical contrast with high imaging speed and resolution. However, PACT
has not been demonstrated to reveal the dynamics of whole cardiac anatomy or vascular system without surgical
procedure (thoracotomy) for tissue penetration. Here, we achieved non-invasive imaging of rat hearts using the
recently developed three-dimensional PACT (3D-PACT) platform, demonstrating the regulated illumination and
detection schemes to reduce the effects of optical attenuation and acoustic distortion through the chest wall; thereby,
enabling unimpeded visualization of the cardiac anatomy and intracardiac hemodynamics following rapidly scanning
the heart within 10 s. We further applied 3D-PACT to reveal distinct cardiac structural and functional changes among
the healthy, hypertensive, and obese rats, with optical contrast to uncover differences in cardiac chamber size, wall
thickness, and hemodynamics. Accordingly, 3D-PACT provides high imaging speed and nonionizing penetration to
capture the whole heart for diagnosing the animal models, holding promises for clinical translation to cardiac imaging
of human neonates.
Introduction
Cardiovascular disease remains the leading cause of
morbidity, responsible for 16% of the world
’
s total
deaths
1
. Diagnostic imaging and therapeutic targets war-
rant preclinical investigations of animal models
2
,
3
to
understand the cardiac disease mechanisms. Over the last
several decades, non-invasive imaging of small animal
models provides in vivo insights into the structural and
functional phenotypes with physiological and clinical
relevance. The current non-invasive imaging modalities
allow from a wide range of vertebrate animal models,
including light-sheet
fl
uorescent microscopy
4
, echo-
cardiography
5
, magnetic resonance imaging (MRI)
6
, x-ray
computed tomography (CT)
7
, positron emission tomo-
graphy (PET)
8
, and/or single-photon emission computed
tomography (SPECT)
9
, and a combination of these ima-
ging techniques
10
–
12
. Each imaging technique provides
the unique degree of tissue penetration, resolution, and
contrast to image the speci
fi
c animal models from zeb-
ra
fi
sh to mouse to swine models, and the combination of
these complementary techniques allows for addressing
the spatial and temporal resolution,
fi
eld of view (FOV),
and relative phenotypes in response to the particular
imaging needs. For example, echocardiography is a por-
table tool to assess cardiac function by interrogating the
contracting/relaxing heart chambers and open/closure of
the valves in real-time. In contrast, CT and MRI are bulky
but offer larger FOV and/or
fi
ner spatial resolution nee-
ded for cardiac anatomy and vascular system
7
,
13
.
Complementary to the well-established imaging mod-
alities, photoacoustic (PA) computed tomography
(PACT) is an emerging imaging technique that combines
functional optical contrast from light illumination and
© The Author(s) 2023
Open Access
This article is licensed under a Creative Commons Attribution 4.0 Internat
ional License, which permits use, sharing, adaptation, distribution and
reproduction
in any medium or format, as long as you give appropriate credit to the origina
l author(s) and the source, provide a li
nktotheCreativeC
ommons license,
and indicate if
changes were made. The images or other third party material in this article are included in the article
’
s Creative Commons license, unless indicated
otherwise in a credit line to the material. If
material is not included in the article
’
s Creative Commons license and your intended use is not permitted by sta
tutory regulation or exceeds the permitted use, you will need to obtain
permission directly from the copyright hol
der. To view a copy of this license, visit
http://creativecommons
.org/licenses/by/4.0/
.
Correspondence: Tzung K. Hsiai (
THsiai@mednet.ucla.edu
)or
Lihong V. Wang (
LVW@caltech.edu
)
1
Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department
of Medical Engineering, Department of Electrical Engineering, California
Institute of Technology, Pasadena, CA, USA
2
Department of Bioengineering, UCLA, Los Angeles, CA, USA
Full list of author information is available at the end of the article
These authors contributed equally: Li Lin, Xin Tong
1234567890():,;
1234567890():,;
1234567890():,;
1234567890():,;
high spatial resolution from acoustic detection. Compared
with echocardiography, PACT offers rich optical contrasts
that are physiologically relevant
14
and suffers a lower level
of speckling artifacts
15
. As compared with MRI, PACT
usually provides a higher imaging speed and a more
portable imaging platform. Unlike X-ray CT, PACT uses
nonionizing illumination to image cardiac vasculature,
avoiding injection of contrast agents. Like PET, PACT can
reveal molecular information
16
with
fi
ner spatial resolu-
tion obviating the need for ionizing radiation.
Despite these unique advantages, the majority of the
published PACT prototypes have not been demonstrated
for non-invasive cardiac imaging of the detailed anato-
mical or functional phenotypes due to suboptimal illu-
mination and detection schemes. This limitation has been
recognized by the previous PACT studies, including
in vivo imaging of mouse hearts with suboptimal clarity
for anatomical/functional details
17
,
18
and ex vivo imaging
of excised/perfused heart with no intracardiac
fl
ow
dynamics
19
,
20
. There are 3 challenges for in vivo cardiac
PACT. (1) Ribs and lungs surrounding the heart partially
block and disturb PA signals. The same limitation exists
in echocardiography; however, PACT only suffers from
one-way acoustic disturbance since the PA signals are
generated in hearts. (2) The high concentration of
hemoglobin and myoglobin renders the myocardium
highly absorptive to light. This property may produce high
PA signal amplitudes; however, it reduces optical pene-
tration into the heart. (3) The periodic heartbeat requires
real-time imaging or motion-correction mechanisms to
give the imager
’
s spatial resolution full play for dynamic
cardiac structures. For example, the time-gating is widely
used in CT
21
, MRI
22
, and PET
23
, dividing every cardiac
cycle into multiple phases and collating all the data from a
speci
fi
c phase for imaging reconstruction.
In this context, we seek to overcome these challenges
and enhance PACT for pre-clinical and physiological
applications. Here, we modi
fi
ed a recently developed
three-dimensional PACT (3D-PACT) platform
24
with
regulated illumination and detection schemes to image
the whole rat heart in vivo. This 3D approach allows for a
hemispherical detection mesh with an extensive view
aperture, detecting PA signals from the heart semi-
panoramically to alleviate the acoustic blockage caused by
ribs and lungs. In addition, the 1064-nm light suffers less
scattering in the biological tissue
25
than the wavelengths
that are commonly used by PACT within the near-
infrared window (680
–
950 nm). During the scanning, we
synchronized the PACT measurement with cardiac cycles
by gating with the electrocardiogram (ECG). The 3D-
PACT scanned the rat heart in 10 s and reconstructed a
series of cardiac images based on a time-gating strategy.
The 3D-PACT along with the ECG-aided synchroniza-
tion largely resolves the challenges from the previous
PACT approaches and reveals the whole rat heart
anatomy non-invasively. Speci
fi
cally, the 3D-PACT sig-
ni
fi
cantly enhances the image clarity and captures the
dynamic changes in the cardiac structure between obese
and control (i.e., lean) rats; namely, chamber size, myo-
cardial wall thickness, and intracardiac
fl
ow. Moreover,
the 3D-PACT is capable of recording the distinct blood
fl
ow in various sizes of the vascular system from healthy,
obese, and hypertensive rats, including the aortas, pul-
monary arteries, left and right coronary arteries. Notably,
the 3D-PACT shows the near-term potential to provide
cardiac image quality comparable to that of MRI
26
, deli-
vering optical contrast for cardiac structure and function,
and providing physiological parameters for blood
fl
ow.
Results
3D-PACT synchronization with ECG-gating for 4D (3D
space
+
time domain) non-invasive cardiac imaging
In the 3D-PACT platform (Fig.
1
a), we applied light
illumination with a pulse repetition rate of 50 Hz to the
subject through an optical window (Figs.
1
a and S1). Four
arc-shaped ultrasonic transducer arrays rotated coaxially
around the optical window for 90° to form a hemi-
spherical detection matrix (Fig. S2 and Methods section).
We covered the imaging aperture with a transparent
membrane to separate the imaging system from the
subject (rat), which was placed in a prone position over
the imaging aperture with chest close to the center
(Fig.
1
b and Methods section). During the 10-s imaging
session, the arrays were scanned for 500 steps (see
Methods section), and the heart was periodically beating
at about 5 Hz
27
. Due to the motion artifacts induced by
the heart beats, the reconstructed heart image was largely
blurred without motion correction (Fig. S3). Therefore,
we applied time-gated motion correction guided by the
synchronized ECG measurement (see Methods). For
example, to reconstruct the heart image when ventricles
are at maximum dilation (end of the ventricular diastole),
we only used the photoacoustic data acquired at the
scanning positions gated to the R waves of the ECG sig-
nals (Fig.
1
c). Thus, the 3D approach reveals the time-
dependent changes in cardiac structure, including the left
vs. right ventricle, free wall vs. septum, pulmonary artery
vs. aortic arch, and left vs. right carotid arteries.
3D-PACT interrogation of rat cardiac anatomy and vascular
system
To address acoustic distortion from the chest ribs and
lung air sacs that encompass the heart, we demonstrate a
large acoustic detection aperture to collect PA signals
from numerous view angles (Fig.
1
d). To further improve
the PACT reconstruction performance of cardiac anat-
omy, we synchronized the 3D-PACT with ECG, scanning
the heart with ~11 steps (i.e., 11 laser pulses) during each
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 2 of 9
a
Imaging aperture
Ultrasonic
array housing
Four arc-shaped
ultrasonic arrays
Optical window
0
0.5
1
Time (s)
-0.2
0
0.2
Voltage (mV)
c
1
0
Norm. PA amp.
Signal from skin
Signal from heart
P
T
d
b
1.76
π
π
solid angle
0.45
π
solid angle
0.22
π
π
solid angle
1
0
Normalized PA amplitude
5 mm
4×256-channel
DAQs
Nose cone
with tooth bar
Air bleeds
Mounting
post
Heating pad
Expanded
laser beam
Heart
Ultrasonic
array
1550
1590
1630
1670
1710
DAQ sampling points
R
Fig. 1 3D-PACT synchronized with ECG for cardiac imaging. a
Representative sketch of the 3D-PACT platform with a close-up view of the
imaging aperture.
b
Experimental setup for the rat heart imaging. The rat was mounted in a prone position, and the expanded light beam was
directed towards the rat
’
s chest from the bottom.
c
Example of the signal synchronization between the 3D-PACT (top) and ECG (bottom)
measurements. The 3D-PACT signal was collected from one transducer element on one array at multiple scanning positions (i.e., on the same
latitude). In the photoacoustic signal diagram, the horizontal axis represents the scanning steps (i.e., time), and the vertical axis stands for the
time-of-
fl
ight signal (data acquisition sampling at 20 MHz). The P, R, and T waves in the ECG signals correspond to atrial contraction, ventricular contraction,
and ventricular relaxation, respectively.
d
The rat heart images reconstructed with different ultrasonic apertures, showing the impact of a sizeable
ultrasonic detection aperture on image reconstruction. The red arcs on the top right corner of each image represent the cross-sectional view of the
ultrasonic detection aperture
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 3 of 9
cardiac cycle. Accordingly, we divided every cardiac cycle
T into 11 phases and reconstructed a volumetric image
for each phase (Fig.
2
a and Supplementary Movie 1).
Since the image of Phase 1 (i.e., Time 0) was recon-
structed using the data acquired at the R waves of the
ECG signals, the heart was close to the start of systole.
During the
1
11
T
6
11
T period, the heart was in systole
but started to relax from
6
11
T. Such a heartbeat pattern
complies with the Wiggers diagram. We further mapped
the
fl
uctuations of PA signals across the heart (Fig. S4) to
take in the heartbeat as dynamic contrast, showing more
spatial movements in the apex of the heart than in the top
regions.
To reveal the internal cardiac structure, we further
showed the cross-sectional images on the sagittal (Fig.
2
b
and Supplementary Movie 2) and coronal planes (Fig.
2
c
and Supplementary Movie 3). In the cross-sectional
images, we can identify the dynamic changes in atria
and ventricles during cardiac contraction (systole) and
relaxation (diastole). In Fig.
2
c, the end ventricular dia-
stolic volume was observed near Time 0, whereas the end
atrial diastolic volume was near
6
11
T. We further encoded
the imaging depth by colors and peeled away the
structures from the chest wall towards the posterior
region of the heart, showing cardiac structures at different
depths (Fig.
2
d). The imaging contrast of the cardiac
anatomy acquired by the 3D-PACT is from optical
absorption of myoglobin from the myocardium and
hemoglobin from the blood
28
.
3D-PACT interrogation of hypertrophic hearts in the obese
rats
It is widely recognized that overweight and obesity are
cardiometabolic risk factors for developing acute coronary
syndromes and stroke
29
. Obesity-associated increase in
cholesterol levels and vascular in
fl
ammation are recog-
nized to cause ventricular remodeling and to reduce blood
perfusion to the myocardium
30
. Here, we demonstrate to
capacity of 3D-PACT to interrogate obesity-induced
cardiac enlargement or known as hypertrophic abnorm-
alities in cardiac anatomy and function by non-invasively
imaging Zucker obese/lean rats of a similar age (Table S1).
Compared with the lean rats, whose cardiac cycles typi-
cally have 11 phases (11 phases/50 Hz
=
0.22 s, i.e., 4.55-Hz
cardiac rate), the three obese rats we imaged have 9
phases for each cardiac cycle (i.e., 5.55-Hz cardiac rate),
Time 0
a
b
c
1
0
Normalized PA amplitude
Depth from the chest wall
d
0 cm
5 mm
1.8 cm
5 mm
IV
PuA
RV
LV
LA
SVC
RA
Aorta
ITV
RA
RV
BA
z = 0 – 1.8 cm
z = 0.5 – 1.8 cm
z = 0.9 – 1.8 cm
Anterior
Posterior
10
11
T
8
11
T
6
11
T
4
11
T
2
11
T
Fig. 2 Rat heart anatomy acquired by the 3D-PACT. a
Front view of the heart within a cardiac cycle. The heart is identi
fi
ed by a magenta circle at
4
11
T. BA, brachiocephalic artery; ITV, internal thoracic vessels; IV, intercostal vessels.
b
Cross-sectional images of the heart on the sagittal plane. Each
image is a maximum amplitude projection (MAP) of a slice marked by the green dashed lines in (
a
) (Time 0).
c
Cross-sectional images of the heart on
the coronal plane. Each image is an MAP of a slice marked by the yellow dashed lines in (
b
) (Time 0). LA, left atrium; LV, left ventricle; PuA, pulmonary
artery; RA, right atrium; RV, right ventricle; SVC, superior vena cava.
d
The same data (at Time 0) shown with color-encoded depths. Shallower
structures were peeled away in the lower images to show the posterior anatomy
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 4 of 9
indicating a higher cardiac rate for augmented cardiac
output to compensate for the increased metabolic
requirements
31
.
The 3D imaging capability enables the measurement of
cardiac anatomy and function. By sectioning the volu-
metric images on the coronal plane (Fig.
3
a, b), we
observed concentric hypertrophy
32
in the obese rat
’
s heart
with increased free wall thicknesses in both left and right
ventricles compared to the lean rat (Fig.
3
c). For thickness
measurement, we selected the middle slice of the heart on
the coronal plane and quanti
fi
ed the full-width-at-half-
maximum of the photoacoustic signals across the wall
section with a length of about 1.3 mm near the middle
(Fig. S5). Ex vivo measurements were performed by
splitting the dissected hearts on the coronal plane near the
middle after imaging, showing concordance with the
in vivo measurements (Fig. S6).
In addition, measurements of the changes in cardiac
volumes enabled us to assess both the systolic and dia-
stolic function. We divided each heart image into ten
sections with speci
fi
c thicknesses (1.04 mm) on the cor-
onal plane and segmented ventricular and atrial regions to
evaluate the volume variations (Fig. S5). To reduce the
individual difference within each group (
n
=
3), we nor-
malized the volumes of ventricles and atria to their values
at Phase 1 (at the end of diastole). The relative changes in
the cardiac function between the lean and obese rats
(Figs.
3
d and S7) were characterized in terms of the end-
systolic ventricular volume and end-diastolic atrial
volume of the cardiac cycle. Probably due to the reduced
strain of ventricles or well-recognized diastolic
dysfunction in the hypertrophic hearts
32
, we observed
slightly lower changes in the average of normalized left
ventricular volumes in obese rats. Similarly, the left atria
in the obese rats developed slightly lower variations in the
normalized volume.
3D-PACT interrogation of cardiovascular hemodynamics
Given PACT
’
s high sensitivity to the hematogenous
contrast, the 3D-approach enabled periodic hemody-
namic interrogation of the cardiovascular system. In a
cardiac image acquired by 3D-PACT (Fig.
4
a), the aorta,
pulmonary artery, right coronary artery, and left coronary
artery were captured. To measure the intravascular
hemodynamics, we selected a region of interest (ROI)
slightly larger than the vessels
’
cross-sections and aver-
aged the photoacoustic signal amplitude within a section
of 10-voxel length (1.3 mm) along the vessel. Since oxy-
hemoglobin dominates the optical absorption at 1064 nm
(i.e.,
μ
a
(oxyhemoglobin)
≈
10 ×
μ
a
(deoxyhemoglobin)
;
μ
a
:
absorption coef
fi
cient)
33
, the changes in spatially averaged
PA signals within the ROI mainly re
fl
ect the variations in
the amount of oxyhemoglobin, which is determined by
the blood volume (or vascular diameter) and oxygen
saturation (sO
2
).
The measurements of hemodynamics in the coronary
arteries introduce distinct myocardial perfusion among
healthy, hypertensive, and obese rats. The averaged PA
signals in the aorta and pulmonary artery, peaked during
ventricular systole (0.4
–
0.5 T), showed slightly higher
variations in hypertensive hearts (blue dashed lines in
Fig.
4
b) than in the control group, yet with insigni
fi
cant
Depth from the chest wall
0 cm
1.8 cm
a
b
5 mm
c
Left ventricle
d
Normalized volume
Right ventricle
Time in one cardiac cycle
RV
LV
LA
RA
Aorta
Lean
Obese
Left ventricle
Left atrium
0.6
2.2
p
<0.01
p
<0.01
0.4
0.8
1.0
1.8
1.4
1.0
00.5TT
1
2
3
1
2
3
Lean
Wall thickness (mm)
Wall thickness (mm)
Normalized volume
Obese
Lean
Obese
Fig. 3 Differences in cardiac anatomy and function between the Zucker obese and lean rats. a
Cardiac anatomy of a Zucker obese rat. Left
panel: Color-encoded depth-resolved image of the heart; middle panel: Cross-sectional image of the heart; Right: Longitudinal sectioning of the
corresponding hypertrophic heart. LA, left atrium; LV, left ventricle; RA, right atrium; RV, right ventricle.
b
Cardiac anatomy of a Zucker lean rat (control
group).
c
Free wall thickness measurements of the left ventricle (top:
p
< 0.01,
n
=
3) and right ventricle (bottom:
p
< 0.01,
n
=
3) between the obese
and lean rats.
d
Relative volume variations of the left ventricles and atriums between the obese and lean rats. The shadow behind curves represents
the standard deviation across multiple rats (
n
=
3)
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 5 of 9
statistical differences. This observation could be sup-
ported by measuring relative volume changes of the left
ventricles (Fig.
4
c). In comparison, we noticed more sig-
ni
fi
cant
fl
uctuations in the relative PA signal amplitudes
from the left coronary arteries in hypertensive hearts
within the cardiac cycle (Fig. S8). We speculate that the
abnormal hemodynamics are related to the accumulated
fat in the damaged arteries, which may affect blood
perfusion
34
.
Similarly, we measured and plotted the hemodynamics
of cardiac vessels in obese rats. Probably due to the aug-
mented cardiac output and obesity-induced arterial
hypertension
35
, we observed higher variations in the
averaged PA signals from the four vessels (red dashed
lines in Fig.
4
b). However, only the right coronary arteries
and aortas in the obese rats showed signi
fi
cant statistical
differences from the control group (Fig. S8). The com-
bined measurements in cardiac anatomy and hemody-
namics depict a potential to provide diagnosis of
cardiovascular diseases in individuals with hypertension
or obesity.
Discussion
For decades, the photoacoustic
fi
eld has faced chal-
lenges in cardiac imaging, including acoustic disturbance
from ribs and lungs, optical attenuation, and motion
artifacts from the periodic heartbeat. Nevertheless, pre-
clinical research and clinical practice are in need of
innovative imaging methods with complementary
advantages to provide high-dimensional physiological
information. In this study, we have demonstrated non-
invasive 3D-PACT of cardiac anatomy and function in
rats, providing substantial improvements in anatomical
image clarity and cardiovascular function measurement.
To reveal the cardiac anatomy, 3D-PACT employs a
hemispherical acoustic detection aperture with large view
angles to mitigate the in
fl
uences induced by ribs and
lungs; The 1064-nm light and uniform illumination also
facilitate deep penetration of the entire heart (Fig.
2
b);
although this study has not provided snapshot or real-
time imaging, the periodic heartbeat allows the ECG-
guided time-gating method to reduce motion-induced
artifacts during a 10-second scan. Clear anatomical ima-
ging further enables measurements of the cardiac func-
tions, including the chamber volume variations and
cardiovascular hemodynamics. For example, the changes
in photoacoustic signals from the cardiac vessels indicate
the variations in blood volume and sO
2
, showing differ-
ences in hemodynamics of the hypertensive, hypertrophic,
and normal control hearts. In summary, the optical ima-
ging contrast in 3D-PACT depicts the distribution of
myoglobin in the myocardium and hemoglobin in the
blood, revealing cardiac anatomy and function with phy-
siological relevance
36
.
While the study was performed based on the mod-
i
fi
cation of a recently published imaging system
24
, addi-
tional technical innovations are critical to producing clear
cardiac images: (1) The ECG-guided time-gating strategy
1
Norm. PA amp.
0
Aorta
Pulmonary
artery
RCA
LCA
Right coronary artery
Aorta
a
c
Pulmonary artery
Left coronary artery
0
0.5T
T
00.5TT
1.0
0.6
1.4
Norm. PA amp.
1.0
0.6
1.4
Norm. PA amp.
b
Control
Hypertensive
Obese
Time in one cardiac cycle
Norm. volume
Time in one cardiac cycle
Time in one cardiac cycle
1.0
0.4
0.6
0.8
0T
0.5T
Fig. 4 Cardiac hemodynamics in control (i.e., healthy), hypertensive, and obese hearts. a
A heart image acquired by the 3D-PACT shows the
aorta, pulmonary artery, right coronary artery (RCA), and left coronary artery (LCA).
b
PA signal
fl
uctuations in the four cardiac vasculatures among
the control, hypertensive, and obese hearts (
n
=
3 for each group).
c
Relative changes in ventricular volume of the control (black solid line),
hypertensive (blue dashed line), and hypertrophic (red dashed line) hearts during a cardiac cycle (
n
=
3 for each group)
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 6 of 9
synchronized the 3D-PACT system with the ECG mea-
surement by sending a spike signal to one of the ECG
’
s
electrodes when the system starts to run (see Methods).
(2) Regulated laser illumination and gated spatial sam-
pling are synchronized to image the heart dynamics
during multiple cardiac cycles (see Methods), providing
an FOV large enough for whole-heart imaging in 3D
space
37
. (3) In PACT, the light illumination, acoustic
detection, and imaging object comprise a comprehensive
signal excitation and detection system. Therefore, animal
positioning that can reliably match the illumination and
detection is usually overlooked but critical for high-
quality imaging (see Methods section).
Based on the improvements in this study, PACT of the
heart can be further developed towards the following
directions: (1) Using ultrasonic arrays with a higher center
frequency (e.g., 5
–
10 MHz) could help reveal detailed
cardiac anatomy in small animals with a higher resolution.
However, the unwanted effects from acoustic attenuation
and rib disturbance will be more signi
fi
cant. (2) To
maintain the lateral FOV with an improved imaging speed,
rearrangement of the ultrasonic transducers (e.g., eight arc
arrays, each with 128 elements) would be implemented to
provide an equally dense azimuthal sampling with less
scanning steps. (3) The data ampli
fi
cation and acquisition
circuits can be modi
fi
ed to integrate ultrasonography,
offering well-interpreted imaging contrast (i.e., informa-
tion) in coregistered images for reference.
3D-PACT is an open imaging platform for preclinical
research and clinical translation by simultaneously pro-
viding cardiac anatomy and function. Compared with
established imaging modalities, PACT of the heart offers
additional physiology-related information with high ima-
ging speed and resolution, yet without the need for
ionizing radiation or invasiveness, making PACT a pro-
mising tool for cardiac imaging of human neonates.
Materials and methods
Optical illumination and acoustic detection
For optical illumination, PACT requires a relatively uni-
form energy deposition within the
fi
eld of view. While for
acoustic detection, high-quality imaging requires a large
view angle (i.e., synthetic aperture) and dense sampling in
time and space. In 3D-PACT, we directed a 1,064 nm laser
beam (DLS9050, Continuum; 50 Hz pulse repetition rate;
5
–
9 ns pulse width) to the optical window mounted on the
scanning axis. The optical window is composed of an
engineered diffuser (EDC-15, RPC Photonics Inc.) and a
condenser lens (ACL25416U-B, Thorlabs Inc.), expanding
the beam to a diameter of ~4.8 cm (Fig. S2). The illumi-
nation area is slightly larger than the imaging object (i.e.,
heart) to lower the requirement of precise animal posi-
tioning. Because the heart is
fi
lled with blood (i.e., high
optical absorption), we used 1064-nm light in this study for
less optical attenuation in biological tissues to achieve deep
penetration. The optical
fl
uenceonthetissuesurface
(~20 mJ/cm
2
at 50 Hz) was limited by the American
National Standards Institutes
’
safety standards
38
.
The acoustic detection module comprises four ultrasonic
arc arrays with a separation of 90 degrees. Each array
has 256 transducer elements with a central frequency of
2.25 MHz, and each element has a dimension of
0.6 × 0.7 mm
2
, generating a divergence angle (i.e., far-
fi
eld
acoustic diffraction) around 60 degrees covering the ima-
ging object semi-panoramically. Here, we applied a hemi-
spherical detection matrix to enlarge the view angle so the
system can acquire acoustic signals propagating in most
independent directions. In a
ddition, the 2.25-MHz center
frequencyofthearrayalsoallowsthedetectionofsuf
fi
-
ciently low frequency signals that suffer less distortion from
the ribs. Notably, this acoustic detection scheme would not
avoid distortions from bones and lungs, but would reduce
their effects on the reconstructed images. Connecting to the
ultrasonic arrays, 1024-channel pre-ampli
fi
cation and
analog-to-digital conversion circuits (PhotoSound, Inc.)
provide one-to-one mapped data acquisition
24
.
ECG-guided time gating and spatial sampling
The data acquisition modules and mechanical scanner
of the 3D-PACT system were triggered by the laser
’
s
“
Lamp Sync Out
”
trigger output. At the same time, the
computer controlled the trigger signal transmission via an
electronic relay in the system
’
s control box. We turned on
the laser and ECG systems shortly before closing the relay,
which coordinated the data acquisition, laser pulses, and
mechanical scanning. We further directed a coaxial cable
from the system
’
s control box to one of the ECG probes
(left or right arm) and output a spike signal much higher
than the ECG spikes when the relay was closed (Fig. S9).
Accordingly, we pinpointed the start time of the photo-
acoustic data acquisition in the ECG signal, achieving the
data synchronization between ECG and 3D-PACT.
We then grouped the PACT data into multiple heart-
beat phases based on the synchronized ECG cardiac cycle
and reconstructed a volumetric heart image for each
phase (Fig.
1
c). Since the heart rate was around 5 Hz and
the pulsed laser illuminated at 50 Hz, each cardiac cycle
can be divided into 9
–
11 phases, enabling the recon-
struction of 9
–
11 volumetric images to depict the beating
heart. The variations in cardiac cycle duration were
caused by differences in weight (e.g., obese versus regular
rats) and anesthesia (i.e., light versus deep anesthesia).
However, the ECG signal shows that the heart rate was
relatively stable during the 10-second scan.
For spatial sampling, each of the four arrays was scan-
ned for 500 steps (50 Hz × 10 s) during the 10-second
imaging, providing 2000 azimuthal sampling positions.
Accordingly, each heart image is reconstructed by ~200
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 7 of 9
(2000/10 phases) azimuthal positions times 256 polar
positions (i.e., each arc array has 256 elements), which
could provide an effective FOV with a lateral diameter of
15 mm
37
and elevational radius of 38 mm.
Animal preparation
Four strains of rats were imaged in this study (Table S1):
(1) Hsd:Sprague Dawley SD rats (120
–
150 g body weight),
(2) Zucker obese rats (200
–
250 g), (3) Zucker lean rats
(120
–
150 g), (4) Spontaneously hypertensive inbred SHR/
NHsd rats (120
–
150 g). We
fi
rst imaged the SD rats to
optimize the experimental setup and demonstrate the
imaging capability (Fig.
2
). The Zucker lean rats were
imaged as the control group to compare the cardiac
anatomy and function with the obese rats (Fig.
3
). We
compared the cardiovascular hemodynamics in the
hypertensive, obese, and SD rats (Fig.
4
).
The imaging aperture was covered by a transparent
membrane (disposable plastic wrap) that separated the rat
from the imaging system and served as a support for the rat.
We then covered a feedback-controlled heating pad with a
zip bag for waterproo
fi
ngandplaceditabovethemem-
brane to maintain the animal
’
s body temperature (Fig.
1
b).
Before imaging, we removed the hair on the chest and
placed the rat above the heating pad. A 3D-printed nose
cone and tooth bar were positioned to provide air with 2%
vaporized iso
fl
urane. During imaging, we placed two cross
lines above the imaging aperture to mark the center posi-
tion of the FOV. We further bent the rat
’
sfrontlimbs
towards the back of the body so that the heart would be
prominent towards the chest wall. After the imaging session
was completed, we euthanized the rat, collected the heart,
and obtained a photo of the dissected heart for reference.
All the animal experiments followed the protocol approved
by the Institutional Animal Care and Use Committee
(IACUC) of the California Institute of Technology.
Image processing and visualization
We used the dual-speed-of-so
und back-projection algo-
rithm
39
implemented in C
++
to reconstruct all images
with mitigated artifacts induced by the acoustic inhomo-
geneity between the biological tissue and the coupling
fl
uid
(D
2
O). Each volumetric image w
as reconstructed with a
voxel size of 0.13 × 0.13 × 0.13 mm
3
. We then processed
each image to improve the contrast through the following
steps: (1) A depth compensation (e
0.81×depth (cm)
)method
was applied to enhance the PA amplitude in deep tissue. (2)
We then used a high-pass
fi
lter to suppress the low-
frequency background. (3) The high-passed image was
further denoised using sparse 4D transform-domain colla-
borative
fi
ltering
40
. (4) To enhance the contrast of blood
vessels, a Hessian-based Frangi vesselness
fi
lter
41
was
applied to the denoised image. (5) Finally, we added the
vesselness-
fi
ltered images (self-norm
alized) with a weighting
factorof0.5backtothehigh-passedimages(witha
weighting factor of 0.5) and obtained the presented images.
The color-encoded images in Figs.
2
and
3
were acquired
using a specialized 3D visualization software package,
namely 3D PA Visualization Studio
42
.
Acknowledgements
This work was sponsored by the United States National Institutes of Health
(NIH) grants R35 CA220436 (Outstanding Investigator Award) and U01
NS099717 (BRAIN Initiative).
Author details
1
Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department
of Medical Engineering, Department of Electrical Engineering, California
Institute of Technology, Pasadena, CA, USA.
2
Department of Bioengineering,
UCLA, Los Angeles, CA, USA.
3
Division of Cardiology, Department of Medicine,
UCLA, Los Angeles, CA, USA.
4
Present address: College of Biomedical
Engineering and Instrument Science, Zhejiang University, Hangzhou, China.
5
Present address: The First Af
fi
liated Hospital, Zhejiang University School of
Medicine, Hangzhou, China
Author contributions
L.V.W. and L.L. conceived and designed the study. L.L. constructed the
hardware system. S.N. and L.L. developed the control program. Y.Z. and R.C.
modi
fi
ed the control program for ECG synchronization. L.L. and X.T. performed
the experiments. L.L., X.T., and Y.Z. analyzed the data. L.V.W., T.K.H., and S.C.
supervised the study. All authors wrote the manuscript.
Data availability
All data are available within the Article and Supplementary Files, or available
from the authors upon request.
Con
fl
ict of interest
All the authors declare no competing interests. L.V.W. has a
fi
nancial interest in
Microphotoacoustics, Inc., CalPACT, LLC, and Union Photoacoustic
Technologies, Ltd., which, however, did not support this work.
Supplementary information
The online version contains supplementary
material available at
https://doi.org/10.1038/s41377-022-01053-7
.
Received: 11 July 2022 Revised: 29 November 2022 Accepted: 1 December
2022
References
1. Word Health Organization.
The Top 10 Causes Of Death
.
https://www.who.int/
news-room/fact-sheets/detail/the-top-10-causes-of-death
(2020).
2. Neeland, I. J. et al. Effects of liragluti
de on visceral and ectopic fat in adults
with overweight and obesity at high cardiovascular risk: a randomised, dou-
ble-blind, placebo-controlled, clinical trial.
Lancet Diabetes Endocrinol.
9
,
595
–
605 (2021).
3. Lobelo, F. et al. Routin
e assessment and promotion of physical activity in
healthcare settings: a scienti
fi
c statement from the American Heart Associa-
tion.
Circulation
137
,e495
–
e522 (2018).
4. Fei, P. et al. Cardiac light-sheet
fl
uorescent microscopy for multi-scale and
rapid imaging of architecture and function.
Sci. Rep.
6
, 22489 (2016).
5. Chetboul, V. Advanced techniques in echocardiography in small animals.
Vet.
Clin. North Am.: Small Anim. Pract.
40
,529
–
543 (2010).
6. Nahrendorf, M. et al. Cardiac magne
tic resonance imag
ing in small animal
models of human heart failure.
Med. Image Anal.
7
, 369
–
375 (2003).
7. Bertolini, G. & Angeloni, L. Vascul
ar and cardiac CT in small animals. in
Com-
puted Tomography-Advanced Applications
(ed Halefo
ğ
lu,A.M.)(Rijeka:Inte-
chOpen, 2017).
8. Wang, Q. et al. Application of animal and human PET in cardiac research.
Am.
J. Cardiovasc. Dis.
8
,24
–
30 (2018).
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 8 of 9
9. Golestani, R. et al. Small-animal SPEC
T and SPECT/CT: application in cardio-
vascular research.
Eur.J.Nucl.Med.Mol.Imaging
37
,1766
–
1777 (2010).
10. Koba,W.,Jelicks,L.A.&Fine,E.J.MicroPET/SPECT/CTimagingofsmallanimal
models of disease.
Am.J.Pathol.
182
,319
–
324 (2013).
11. Nanni, C. & Torigian, D. A. Applications of small animal imaging with PET, PET/
CT, and PET/MR imaging.
PET Clin.
3
, 243
–
250 (2008).
12. Izaguirre, E. W. et al. Dual modality micro-SPECT and micro-CT for small animal
imaging: technical adv
ances and challenges.
Proceedings of the SPIE 5923,
Penetrating Radiation Systems and Applications VII
.(SPIE,2005).
13. Botnar, R. M. & Makowski, M. R. Cardio
vascular magnetic resonance imaging in
small animals.
Prog.Mol.Biol.Transl.Sci.
105
,227
–
261 (2012).
14. Yao, J. J., Xia, J. & Wang, L. V. Multiscale functional and molecular photo-
acoustic tomography.
Ultrasonic Imaging
38
,44
–
62 (2016).
15. Guo,Z.J.,Li,L.&Wang,L.V.Onthespeckle-freenatureofphotoacoustic
tomography.
Med. Phys.
36
, 4084
–
4088 (2009).
16. Weber, J., Beard, P. C. & Bohndiek, S. E. Contrast agents for molecular pho-
toacoustic imaging.
Nat. Methods
13
, 639
–
650 (2016).
17. Deán-Ben, X. L. et al. Spiral volumetric optoacoustic tomography visualizes
multi-scale dynamics in mice.
Light Sci. Appl.
6
, e16247 (2017).
18. Ivankovic, I. et al. Volumetric optoac
oustic tomography enables non-invasive
in vivo characterization of impaired he
art function in hyp
oxic conditions.
Sci.
Rep.
9
, 8369 (2019).
19. Lin, H. C. A. et al. Ultrafast volumetric optoacoustic imaging of whole isolated
beating mouse heart.
Sci. Rep.
8
, 14132 (2018).
20. Özsoy, Ç. et al. Ultrafast four-dimensional imaging of cardiac mechanical wave
propagation with sparse optoacoustic sensing.
Proc. Natl Acad. Sci. USA
118
,
e2103979118 (2021).
21. Desjardins, B. & Kazerooni, E. A. ECG-gated cardiac CT.
Am.J.Roentgenol.
182
,
993
–
1010 (2004).
22. Nacif,M.S.etal.Cardiacmagneti
c resonance imaging and its electro-
cardiographs (ECG): tips and tricks.
Int. J. Cardiovasc. Imaging
28
, 1465
–
1475
(2012).
23. Lassen, M. L., Kwiecinski, J. & Slomk
a, P. J. Gating approaches in cardiac PET
imaging.
Pet. Clin.
14
, 271
–
279 (2019).
24. Lin, L. et al. High-speed three-dime
nsional photoacoustic computed tomo-
graphy for preclinical resear
ch and clinical translation.
Nat. Commun.
12
,882
(2021).
25. Jacques, S. L. Optical properti
es of biological tissues: a review.
Phys. Med. Biol.
58
,R37
–
R61 (2013).
26. Price, A. N. et al. Cardiovascular magnetic resonance imaging in experimental
models.
Open Cardiovasc. Med. J.
4
,278
–
292 (2010).
27. Azar,T.,Sharp,J.&Lawson,D.Heart
rates of male and female sprague-dawley
and spontaneously hypertensive rats housed singly or in groups.
J. Am. Assoc.
Lab. Anim. Sci.
50
,175
–
184 (2011).
28. Lin, L. et al. In vivo photoacoustic tomography of myoglobin oxygen
saturation.
J. Biomed. Opt.
21
, 061002 (2015).
29. Piché, M. E., Tchernof, A. & Després, J. P. Obesity phenotypes, diabetes, and
cardiovascular diseases.
Circ. Res.
126
,1477
–
1500 (2020).
30. Clark, L. T. Cholesterol and heart disease: current concepts in pathogenesis
and treatment.
J. Natl Med. Assoc.
78
,743
–
751 (1986).
31. Alpert, M. A. & Hashimi, M. W. Obesity and the heart.
Am.J.Med.Sci.
306
,
117
–
123 (1993).
32. Aurigemma, G. P., De Simone, G. & Fitzgibbons, T. P. Cardiac remodeling in
obesity.
Circ.: Cardiovasc. Imag.
6
,142
–
152 (2013).
33. Pantaleone, C.
Cancer Immunotherapy and Biological Cancer Treatments
(ed.
Arnouk, H.) (IntechOpen, 2019).
34. Mayet, J. & Hughes, A. Cardiac and vascular pathophysiology in hypertension.
Heart
89
,1104
–
1109 (2003).
35. Artham, S. M. et al. Obesity and hyperte
nsion, heart failure, and coronary heart
disease-risk factor, paradox, and recommendations for weight loss.
Ochsner J.
9
,124
–
132 (2009).
36. Hendgen-Cotta, U. B., Kelm, M. & Rassaf, T. Myoglobin functions in the heart.
Free Radic. Biol. Med.
73
,252
–
259 (2014).
37. Hu, P. et al. Spatiotemporal antialiasing in photoacoustic computed tomo-
graphy.
IEEE Trans. Med. Imaging
39
,3535
–
3547 (2020).
38. American National Standards Institute.
American National Standard for the Safe
Use of Lasers ANSI, z136
. 131-2014 (Laser Institute of America, 2014).
39. Li, L. et al. Single-impulse panoramic
photoacoustic computed tomography of
small-animal whole-body dynamics at
high spatiotemporal resolution.
Nat.
Biomed. Eng.
1
, 0071 (2017).
40. Maggioni, M. et al. Nonlocal transform-domain
fi
lter for volumetric data
denoising and reconstruction.
IEEE Trans. Image Process.
22
,119
–
133 (2013).
41. Frangi, A. F. et al. Multiscale vessel enhancement
fi
ltering.
Proceedings
of the First International Conference on Medical Image Computing and
Computer-Assiste
d Intervention.
(pp. 130
–
137. (Springer, Cambridge,
1998).
42. Cho,S.etal.3DPHOVIS:3Dphoto
acoustic visualization studio.
Photoacoustics
18
, 100168 (2020).
Lin et al.
Light: Science & Applications
(2023) 12:12
Page 9 of 9