The dynamic compressive behavior of beryllium bearing bulk
metallic glasses
H. A. Bruck
Idaho National Engineering Labs, Mail Stop 2218, Idaho Falls, Idaho 83415-2218
A. J. Rosakis
Department of Aeronautics, California Institute of Technology, Mail Stop 105-50,
Pasadena, California 91125
W. L. Johnson
Department of Materials Science, California Institute of Technology, Mail Stop 105-50,
Pasadena, California 91125
(Received 14 March 1995; accepted 26 October 1995)
In 1993, a new beryllium bearing bulk metallic glass with the nominal composition
of Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
was discovered at Caltech. This metallic glass can be
cast as cylindrical rods as large as 16 mm in diameter, which permitted specimens to be
fabricated with geometries suitable for dynamic testing. For the first time, the dynamic
compressive yield behavior of a metallic glass was characterized at strain rates of 10
2
to
10
4
y
s by using the split Hopkinson pressure bar. A high-speed infrared thermal detector
was also used to determine if adiabatic heating occurred during dynamic deformation of
the metallic glass. From these tests it appears that the yield stress of the metallic glass is
insensitive to strain rate and no adiabatic heating occurs before yielding.
I. INTRODUCTION
Many components formed from high strength alloys
experience dynamic mechanical loading during their
fabrication and use, which cause the materials to deform
at very high strain rates. In order to understand the limita-
tions of using high strength alloys in these processes, it is
necessary to quantify the material’s dynamic constitutive
behavior. Insight into this behavior has been customarily
obtained using Taylor anvil tests
1
or Kolsky bars.
2
One class of high strength alloys that can be poten-
tially used in many commercial and industrial applica-
tions is known as amorphous metal alloys or metallic
glasses. Previously, cooling rate requirements have lim-
ited the size of metallic glass specimens available for
mechanical testing to cylindrical rods no larger than
2 mm in diameter. Larger specimens are required to
obtain reliable data on the mechanical behavior of these
alloys. Recently, metallic glasses containing beryllium
have been fabricated as ingots up to 16 mm in diame-
ter by directly quenching from the melt.
3
Mechanical
tests have been conducted on specimens manufactured
from these beryllium bearing bulk metallic glasses.
4
The
quasistatic constitutive behavior of these bulk metallic
glasses was characterized in these tests.
At that time, all of the constitutive characterization
of metallic glasses was performed in quasistatic tests. No
data are currently available on the dynamic constitutive
behavior of metallic glasses. An attempt was made
by Cline and Reaugh
5
to determine the dynamic yield
strength of a palladium alloy metallic glass using the
Taylor anvil test; however, they concluded that this was
not appropriate.
The goal of this work is to characterize the dynamic
compressive yield behavior of the Zr
41.25
Ti
13.75
Cu
12.5
-
Ni
10
Be
22.5
metallic glass using a Kolsky bar. Ther-
mal measurements were also made using a high speed
infrared temperature sensor
6
to investigate whether adia-
batic heating occurs during the dynamic deformation of
the glass.
II. KOLSKY BAR THEORY
The basic principle of the Kolsky bar is illustrated
in Fig. 1(a). The setup basically consists of two elastic
pressure bars,
AB
known as the incident or input bar and
CD
known as the transmitter or output bar. A specimen
is placed between the two bars. A striker bar impacts
the input bar at
A
. This sets up a pressure pulse within
the incident bar with a strain
e
I
s
t
d
. When this pulse
reaches
B
, the end of the incident bar, part of the wave
is reflected with a strain
e
R
s
t
d
and part of the wave is
transmitted to the output bar with a strain
e
T
s
t
d
. Strain
gauges are placed in the middle of the elastic pressure
bars to provide the time-resolved signals of the incident,
reflected, and transmitted pulses. The pulse is reflected
from the interface of the bar and specimen at
B
due to
impedance mismatch.
When the specimen is deforming uniformly, the
strain rate
Ÿ
e
within the specimen is proportional to the
J. Mater. Res., Vol. 11, No. 2, Feb 1996
1996 Materials Research Society
503
Help
Comments
Welcome
Journal of
MATERIALS RESEARCH
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
(a)
(b)
FIG. 1. (a) Kolsky bar used for dynamic compression tests. ( b) Stress
and strain analysis of homogeneously deformed specimen.
reflected wave’s amplitude. Similarly, the stress within
the sample is proportional to the amplitude of the trans-
mitted wave. This can be shown by the following analy-
sis [Fig. 1( b)]. In this figure,
A
is the cross-sectional area
of the elastic bars and
A
0
is the cross-sectional area of the
specimen. The strain rate in the deforming specimen is
d
e
dt
≠
Ÿ
e
s
t
d
≠
y
1
s
t
d
2y
2
s
t
d
L
,
(1)
where
y
1
and
y
2
are particle velocities at
B
and
C
,
respectively, and
L
is the length of the specimen. From
characteristic relations for one-dimensional wave propa-
gation, assuming no wave dispersion occurs, one has
y
1
≠
c
0
s
e
I
2e
R
d
,
(2a)
y
2
≠
c
0
e
T
,
(2b)
where
c
0
≠
p
E
y
r
is the longitudinal bar wave speed for
the elastic bars,
E
is Young’s modulus of the elastic bars,
and
r
is the density of the elastic bars. From Eqs. (1)
and (2):
Ÿ
e
s
t
d
≠
c
0
L
f
e
1
s
t
d
2e
R
s
t
d
2e
T
s
t
dg
.
(3)
The average stress in the specimen is
s
s
t
d
≠
P
1
s
t
d
1
P
2
s
t
d
2
A
0
,
(4)
where
P
1
and
P
2
are the forces at the incident
bar/specimen and specimen/transmitter bar interfaces,
respectively. Assuming that no wave dispersion occurs:
P
1
≠
EA
0
f
e
I
s
t
d
1e
R
s
t
dg
,
(5a)
P
2
≠
EA
0
e
T
s
t
d
.
(5b)
From Eqs. (4) and (5):
s
s
t
d
≠
EA
0
2
A
f
e
I
s
t
d
1e
R
s
t
d
1e
T
s
t
dg
.
(6)
When
the
specimen
is
deforming
uniformly
(homogeneously), the stress at the incident bar/specimen
interface equals the stress at the transmitter bar/specimen
interface. From Eq. (5):
P
1
s
t
d
≠
P
2
s
t
d
,
(7)
which implies that:
e
I
s
t
d
1e
R
s
t
d
≠
e
T
s
t
d
.
(8)
From Eqs. (3), (6), and (8):
s
s
t
d
≠
EA
0
A
e
T
s
t
d
,
(9)
Ÿ
e
s
t
d
≠
2
2
c
0
L
e
R
s
t
d
.
(10)
Integrating Eq. (10), we obtain the time history of the
strain as follows:
e
s
t
d
≠
Z
t
0
Ÿ
e
s
t
d
d
t
.
(11)
The stress-strain behavior is completely determined by
measuring
e
T
s
t
d
and
e
R
s
t
d
on the elastic pressure bars
using strain gauges.
III. HIGH-SPEED INFRARED THERMAL
MEASUREMENTS
High-speed, noncontact techniques have been devel-
oped for measuring temperature fields on the surfaces
of deforming objects.
6,7
These techniques use detectors
to measure the infrared thermal signatures generated by
the deforming object. An Offner imaging system first
employed by Zehnder and Rosakis
6
was used for the
dynamic compression tests to focus the surface of the
compression specimens onto a single infrared detector.
The Offner system has a fixed 1 : 1 magnification, so that
optics were used with a focal length of approximately
160 mm and an object distance of 320 mm in order to
protect the optics by keeping them a sufficient distance
away from the test specimen.
A single EG&G J15D12 HgCdTe infrared detector
with dimensions 100
m
m square and a temperature range
of approximately 300 to 1000 K was used to detect
the temperature. The HgCdTe detector was cooled with
liquid nitrogen, and the voltage from the detector was
outputted to a preamp with a bandpass of 5 Hz to 1 MHz,
well beyond the detector’s cutoff frequency of 400 kHz.
Because the sensor had a very high signal-to-noise ration,
the preamp was connected to an amplifier set at
20
3
and
a filter set at 10 kHz, which in turn was connected to
a Nicolet 440 digital oscilloscope that was connected to
the strain gauges on the Kolsky bars.
504
J. Mater. Res., Vol. 11, No. 2, Feb 1996
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
The HgCdTe detector integrates wavelengths rang-
ing from 2
m
mto12
m
m. The voltage produced by the
detectors is related to the energy emitted by the specimen
through an integral, depending upon the emissivity of
the material,
e
s
l
,
T
d
, the spectral responsivity of the
HgCdTe detector,
R
s
l
d
, and the black body radiation
P
s
l
d
, as follows
7
:
y
s
T
,
T
0
d
≠
AA
D
b
3
Z
15
m
m
2
m
m
R
s
l
d
f
P
s
l
,
T
d
e
s
l
,
T
d
2
P
s
l
,
T
0
d
e
s
l
,
T
0
d
g
d
l
,
(12)
where the integration variable
l
is the wavelength of
the radiation,
A
is the amplification,
A
D
is the detector
area, and
b
is the fraction of energy transmitted to the
detectors by the optical system (related to the aperture).
When evaluated for HgCdTe detectors, this relationship
is nearly linear on a log-log plot of voltage versus
temperatures. In order to determine this relationship for
the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
metallic glass, a 1.5 mm
thick plate specimen with a thermocouple clamped to
its surface was heated in an oven. The voltage output
from the detectors and the thermocouple readings were
then recorded as the specimen cooled. This calibration
datum was then plotted on a log-log plot, and the linear
relationship was determined (Fig. 2). This relationship
was then used to convert the voltage record from the
detector into a temperature measurement.
IV. EXPERIMENTAL PROCEDURE
Kolsky bar specimens were fabricated as cylindrical
rods with the nominal composition Zr
41.25
Ti
13.75
Cu
12.5
-
Ni
10
Be
22.5
using commercially pure (
.
99.5%
) elements.
The rods were cast by injecting the molten alloy into a
copper mold with cylindrical cavities that are 2.5 mm
in diameter and 30 mm long. This geometry was chosen
because of the ease with which it could be fabricated
FIG. 2. Thermal calibration curve for Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
.
and the significant ductility it exhibited in quasistatic
compression tests.
4
Details of the alloy design and
casting procedure can be found in Ref. 3. Periodically,
disks of approximately 250
m
m in thickness were ma-
chined from the rods and placed in an Inel CPS-120
x-ray diffractometer to verify that the structure was
amorphous.
After casting the rods, Kolsky bar specimens were
machined with length to diameter
s
l
y
d
d
aspect ratios
ranging from 1 : 2 to 2 : 1. The ends of the specimens
were then mechanically polished with 600 grit grinding
paper using a V-block to ensure flat and parallel surfaces.
By measuring the length of the specimen at four different
locations along the specimen’s edge with a Mitutoyo
micrometer, the surfaces were found to be flat and
parallel to within 3
m
m, the resolution of the micrometer.
The dynamic testing of the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
-
Be
22.5
metallic glass was conducted using a split Hop-
kinson bar with elastic bars made from C350 maraging
steel that has a yield stress of 2.5 GPa. The bars were
1.22 meters in length and 1.27 cm in diameter. Speci-
mens were placed between the elastic bars using a pair
of WC insets coated with “Molygraph Extreme Pressure
Multi-purpose” grease to reduce end effects. The WC
inserts were machined to a diameter of approximately
0.85 cm in order to impedance match them with the
C350 maraging steel. A striker bar was used that was
manufactured from the same material as the elastic
bars and with the same diameter, but only 10 cm in
length. Strain gauges from Micromeasurements Group
were placed midway from the ends of the elastic bars
to measure the strain response on the surface of the
bar from the propagating waves. The strain gauges
were attached to quarter Wheatstone bridges whose
voltage outputs were measured with a Nicolet 440 digital
oscilloscope using a sampling rate of 5 MHz.
Sometimes when the striker bar impacts the incident
bar, there is some misalignment which can cause a
variation in the amplitude of the incident stress pulse
often referred to as “ringing.” Consequently, there is
some dispersion of the stress pulse that is detected by
the strain gauges well after the stress pulse has passed
underneath the strain gage. This makes it extremely
difficult to detect the start of the reflected wave when
it passes through the same strain gage.
In order to reduce dispersion from the incident stress
pulse, a copper insert was placed between the striker
bar and the incident bar as described by Follansbee in
the
Metals Handbook.
8
The resulting stress pulse gen-
erated by using the copper inserts was a ramp function
instead of the step function generated when the copper
inserts are removed. The period of the stress pulse was
nearly quadrupled by the presence of the copper inserts.
Theoretically, the length of the stress pulse would be
twice the length of the striker bar, which implies that
J. Mater. Res., Vol. 11, No. 2, Feb 1996
505
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
the stress pulse generated using copper inserts would be
eight times the length of the striker bar. A constraint on
the length of the stress pulse is it should be less than the
length of the bar if there is to be no overlap of incident
and reflected signals underneath the strain gage when
it is positioned midway from the ends of the bar. So,
the striker bar should be less than 15 cm in length if the
stress pulse is to be less than the length of the elastic bar,
which dictated that the 10 cm long striker bar be used.
Striker bar velocities varied from 5 m
y
sto20m
y
s,
which generated strain rates in the specimen of 300
to 6000
y
s, depending on the length of the specimen.
One problem with using the copper inserts was the
stress pulses generated by their deformations were of
insufficient amplitude to obtain strain rates in the speci-
men greater than about 2000
y
s to 3000
y
s, depending
on the length of the specimen. In order to obtain the
higher strain rates, it was necessary to remove the copper
inserts.
V. EXPERIMENTAL RESULTS AND DISCUSSION
Strain gage data and the thermal measurements from
a typical Kolsky bar test on the Zr
41.25
Ti
13.75
Cu
12.5
-
Ni
10
Be
22.5
metallic glass can be seen in Figs. 3 and 4.
The dynamic stress-strain curve and strain-strain rate
curve reduced from these signals can be seen in Figs. 5
and 6. Yield stresses were obtained for a variety of
strain rates and plotted in Fig. 7. Catastrophic failure of
dynamic compression specimens occurred due to shear
banding along a plane oriented 45
±
to the loading axis as
seen in Fig. 8, which is identical to the failure observed
in quasistatic tests.
Thermal measurements indicated that temperatures
varied dramatically after the specimen failed, reaching
peak values of approximately 500
±
C. Since the speci-
men failed by shear binding, it is possible that the
variations in the thermal trace are due to the shear band
passing through the detector and that the peak values
correspond to the maximum temperature reached within
FIG. 3. Strain gage signals from dynamic compression test.
FIG. 4. Temperature and loading recorded from dynamic compres-
sion test.
FIG. 5. Stress-strain curve reduced from strain gage signals in Fig. 5.
the shear band after deformation. This temperature is ap-
proaching the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
metallic glass
melting point
3
of 663
±
C, which would imply that the
deformation within the shear band is so intense that
the material within the band could be melting. It has
also been observed that there is a temperature of 100
±
C
which is sustained for a period of time after the specimen
has yielded. Since the homogeneous plastic deformation
is too limited to account for this temperature rise, it could
be due to the formation of micro-shear bands.
This is the first time transient thermal measurements
have been made on a metallic glass during a shear band-
ing event. However, more definite conclusions could be
made about the origin of the thermal measurements if an
array of detectors had been used to detect the inhomoge-
neous thermal distribution that would be associated with
shear banding.
Evidence that localized melting may be responsible
for the observed thermal measurements was obtained
from a scanning electron micrograph of the failure
surface, which can be seen in Fig 9. The only mor-
phology that appears on the failure surface are vein-like
506
J. Mater. Res., Vol. 11, No. 2, Feb 1996
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
FIG. 6. Strain-strain rate curve reduced from strain gage signals
in Fig. 5.
patterns. These features are identical to those observed
on the failure surfaces of other metallic glass specimens
tested in compression. An explanation for these features
has been proposed which asserts that the viscosity of
the material within a shear band is reduced consid-
erably because of adiabatic heating generated by the
enormous plastic deformations within the shear band
which preceded catastrophic failure.
9
As the material
separates, pores are opened up, leaving behind the vein-
like morphology. This hypothesis has been confirmed by
shear experiments performed on a viscous material that
has been placed between glass slides.
10
After the slides
are sheared apart, a vein-like morphology is revealed on
the separated surfaces.
The dynamic compressive stress-strain response in
Fig. 5 is very similar to the quasistatic compressive
stress-strain response reported in Ref. 4. For the sake
of comparison, the quasistatic compressive stress strain
curve is shown in Fig. 10. The material exhibits an
almost textbook case elastic-perfectly plastic response
with yielding occurring at approximately 2% elastic
FIG. 7. Yield stress versus strain rate from the dynamic compres-
sion tests.
FIG. 8. Orientation of failure surface for 2.5 mm dynamic compres-
sion test specimen.
strain. From Fig. 7, it appears that the dynamic com-
pressive yield stress is within 5% of the quasistatic
value of 1.89 GPa reported in Ref. 4 for strain rates
below 1500
y
s. However, for strain rates greater than
3000
y
s, the compressive yield stress
appears
to decrease
monotonically with strain rate (strain rate softening). At
first this was believed to have been due to adiabatic
heating of the specimen; however, thermal measurements
indicated that no heating occurred prior to the onset of
yielding (Fig. 4).
During the dynamic compression tests, it was ob-
served that the WC inserts were failing in tests conducted
at strain rates above 3000
y
s. If the WC inserts were fail-
ing before the specimens failed, then the observed strain
rate softening could be due to the failure of the inserts.
To test this theory, the WC inserts were removed and the
compression tests repeated. By removing the inserts, the
FIG. 9. Scanning electron micrograph of compressive failure surface.
J. Mater. Res., Vol. 11, No. 2, Feb 1996
507
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
FIG. 10. Stress-strain curve from quasistate compression test.
stress concentrations at the end of the specimen might
not be reduced which may cause permanent failure. The
results from these tests in Fig. 7 indicate that removing
the WC inserts did not alter the observed yield stress of
the specimens at strain rates below 1500
y
s, while still
exhibiting the same softening behavior above 3000
y
s.
Therefore, it was decided that the WC inserts should be
removed from the dynamic compression tests since they
were not affecting the test results.
No other observations were made during the dy-
namic compression tests that might explain the apparent
strain rate softening. Therefore, it became necessary to
reassess the conditions under which the dynamic com-
pression tests were performed to rule out any possible
artifacts in the testing system that might be responsible
for the apparent strain rate softening.
There are three basic, underlying assumptions used
in the reduction of split Hopkinson pressure bar test data.
They are as follows: (i) The specimen undergoes ho-
mogeneous deformation during the experiment. (ii) The
various stress pulses (e.g., reflected and transmitted)
encountered in the split Hopkinson bar undergo minimal
dispersion. (iii) The bars remain elastic at all times, and
the ends of the bars in contact with the specimen remain
flat and parallel throughout the experiment.
For the strain rates above 3000
y
s, the magnitude
of the incident stress pulse was 400 MPa, well below
the 2.5 GPa yield stress of the elastic bars. Therefore,
assumption (iii) was not violated. Verifying that the
other two assumptions were not violated required using
analyses developed by Ravichandran and Ghatuparthi.
11
VI. LIMITING STRAIN RATE FOR A METALLIC
GLASS SPECIMEN IN A KOLSKY BAR
Ravichandran used a simple analysis to establish
the conditions under which a specimen must be tested
to obtain homogeneous deformations consistent with
assumption (i).
11
This analysis assumed that there must
be some minimum time which was required for the
waves reflected back and forth inside a specimen to
equilibrate. For a specimen which exhibits a linearly
elastic material response up to some failure strain
e
f
,
the limiting strain rate
Ÿ
e
l
was derived to be:
Ÿ
e
l
≠
e
f
c
a
L
,
(13)
where
c
is the specimen’s longitudinal bar wave velocity,
L
is the length of the specimen, and
a
is a nondimen-
sional parameter whose value depends only on the shape
of the incident pulse.
From an analysis of one-dimensional wave propaga-
tion in the elastic bars and specimen using the method
of characteristics,
11
it was determined that for typical
incident pulses the gradient of stress along the length of
the specimen was within 5% of the mean stress in the
specimen when
a
≠
4
. This meant that for the metallic
glass specimens used for the dynamic compression test,
which had lengths of 5 mm or less and a longitudinal bar
wave velocity of 3925 m
y
s determined from ultrasonic
measurements,
4
the limiting strain rates were 3925
y
s
when the failure strain was at least 2%. Although this
value is larger than the 3000
y
s strain rate at which the
softening began to occur, it was still close enough that it
was decided to repeat the tests using specimens with
aspect ratios of 0.7 to 1, which would increase the
limiting strain rate to above 8000
y
s and eliminate any
questions about the homogeneity assumptions.
It has already been observed quasistatically that by
lowering the aspect ratio the yield stress increases be-
cause a plain strain stress state is approached. Therefore,
the conditions appeared encouraging for these tests. In
Fig. 7, it can be seen that the yield stress results from
these tests were identical to the test results for the longer
specimens. Thus, it was concluded that the plane strain
stress state was not being approached as the aspect ratios
of the test specimens were reduced below 1 and that the
inhomogeneity of the stress field was not responsible for
the apparent softening behavior.
VII. EFFECTS OF DISPERSION IN A KOLSKY
BAR AT HIGH STRAIN RATES
Ceramics have exhibited the same apparent strain
rate softening behavior as the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
-
Be
22.5
metallic glass at strain rates above 2700
y
s.
12
Dispersion effects were also analyzed by Ravichandran
and Ghatuparthi to determine if they would account for
the apparent softening behavior of ceramics.
11
Because
of dispersion effects, the amplitude and duration of a
stress pulse changes as it propagates in a bar. The extent
of geometrical dispersion depends on the dominant fre-
quency components contained in the pulse that is being
508
J. Mater. Res., Vol. 11, No. 2, Feb 1996
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
propagated, which is determined by the duration and
shape of the pulse.
13,14
From Ravichandran’s analysis, the effects of dis-
persion are minimized when the dominant frequency
component in the frequency spectrum of the input pulse
in a cylindrical bar is such that
a
y
L,
0.1
. Here,
a
is
the radius of the bar and
L
is the wavelength associated
with the dominant frequency component. This condition
can be rewritten as:
√
c
0
c
p
!
v
v
0
,
0.1 ,
(14)
where
v
0
is the fundamental frequency associated with
the bar,
v
is the fundamental frequency of the pulse,
c
p
is the characteristic propagation velocity for the given
frequency, and
c
0
is the longitudinal bar wave velocity.
The fundamental frequency associated with the bar is
given by:
v
0
≠
2
p
μ
c
0
a
∂
.
(15)
The value of
c
p
for large wavelengths can be obtained
from the Rayleigh approximation
15
:
c
p
c
0
≠
1
2n
2
μ
a
L
∂
2
,
(16)
where
n
is Poisson’s ratio.
For the Kolsky bars used in the dynamic compres-
sion experiments,
a
≠
6.35
mm,
n
≠
0.3
, and
c
0
≠
4970
m
y
s. Thus, plugging these values into Eq. (15)
yields
v
0
≠
4.92
3
10
6
rad
y
s. Using the value for
c
p
at
a
y
L
≠
0.1
from Eq. (16), a limiting value of
4.88
3
10
5
rad
y
s for the fundamental frequency of the pulse
s
v
d
is obtained. The corresponding period of the pulse,
T
,
can be calculated as follows:
v
0
≠
2
p
T
.
(17)
Assuming that the time to failure coincides with
the peak of the input pulse, and from symmetry, the time
to failure
t
f
will be
T
y
2
. The minimum time to failure
t
f
should be at least 6.44
m
s to minimize dispersion.
Using the failure strain of 2% for the Zr
41.25
Ti
13.75
-
Cu
12.5
Ni
10
Be
22.5
metallic glass, the limiting strain rate
would be 3100
y
s. This is approximately the strain rate
at which the apparent strain rate softening begins. Since
the reflected pulse is also affected by dispersion, it is
not unreasonable to believe that the strain rates, like the
stresses, are also underestimated. So, the true strain rates
are probably much higher than 3000
y
s, which would
mean that the strain rate softening occurs well beyond
the calculated limiting strain rate.
There are two methods to correct the dispersion
problem. The first is to directly correct for the dispersion
in the signal.
15
This requires that a sufficient number
of data points be sampled so that the contribution of
each significant frequency component can be extracted
by means of a Fourier transform. At the sampling
rates used in the dynamic compression experiments,
approximately 80 data points could be obtained for strain
rates above 3000
y
s. Unfortunately, this is not enough
data to characterized properly the signal in terms of its
significant frequency components. Therefore, dispersion
in the signal cannot be corrected directly.
The second method involves placing a strain gage
on the surface of the specimen to measure directly the
specimen strain.
12
In order to interpret the corresponding
stresses, a linearly elastic material response must be as-
sumed using a Young’s modulus
4
of 93 GPa to compute
the stresses. This assumption appears to be reasonable
given that the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
metallic glass
exhibits linearly elastic behavior up to yield and that the
ductility of the specimen is very limited.
A strain gage from JP Technologies was placed on
the surface of a dynamic compression test specimen,
and tests were repeated for strain rates above 3000
y
s
and at approximately 230
y
s. For strain rates of 230
y
s,
the specimen did not fail and the deformation was
purely elastic, as evidenced by a stress-strain response
identical to the linear portion of the curve in Fig. 5. The
strains were recorded from the specimen strain gage and
compared with the strain derived from the transmitted
pulse, assuming a linearly elastic response (Fig. 11).
From these low strain rate results, it appears that the
strains from the specimen strain gage are within 5% of
the strains derived from the transmitted pulse with the
greatest deviation occurring at the peak strain level. This
is not unexpected since strain gage signals are rarely
accurate to within 10% of the actual strain depending
on the strength of the adhesion between the gage and
the specimen surface. The strain rate was computed by
curve fitting the linear portion of the strain signal from
the specimen strain gage.
FIG. 11. Dynamic strain response from the specimen strain gage and
derived from the transmitted signal for a strain rate of 230
y
s.
J. Mater. Res., Vol. 11, No. 2, Feb 1996
509
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
After elastically loading the specimen at low strain
rates, the specimen was reloaded at a strain rate above
3000
y
s. A typical result can be seen in Fig. 12 where the
strain rate was determined to be approximately 3400
y
s.
From these results it is obvious that the strain derived
from the transmitted signal does not correlate with the
strain from the specimen strain gage. In fact, the strain
from the strain gage has a bandwidth of approximately
10
m
s, while the strain derived from the transmitted
signal has a bandwidth of about 14
m
s. This difference
in bandwidths was also observed in high strain rate
compression tests conducted on ceramics and attributed
to dispersion effects.
The strain response from the specimen strain gage
shows a dramatic increase in strain rate at approximately
2.0% strain, which is most likely due to the initiation of
shear banding. Therefore, this was taken to be the yield
strain of the specimen. The peak strain derived from the
transmitted bar is approximately 1.3%, a value 35% be-
low the yield strain. The differences in peak strain were
also observed in the high strain rate experiments con-
ducted on ceramics and attributed to dispersion effects.
The yield stress was therefore calculated to be
approximately 1.9 GPa, which is within 5% of the value
derived from the transmitted signal at strain rates below
3000
y
s. This yield stress can be seen in Fig. 13 plotted
against the data from Fig. 7 along with another data point
obtained using a strain gage on a specimen tested at a
strain rate of approximately 4400
y
s that had not been
prestrained. The data points obtained using specimen
strain gauges are larger than any of the yield stresses
derived from the transmitted signal at strain rates above
3000
y
s, clearly indicating that dispersion effects are
responsible for the apparent softening behavior observed
at strain rates above 3000
y
s.
Although peak strains of 3.5% were measured from
the specimen strain gage, it is difficult to interpret the
corresponding load levels. Obviously, if shear banding
FIG. 12. Dynamic strain response from the specimen strain gage and
derived from the transmitted signal for a strain rate of 3400
y
s.
FIG. 13. Data from Fig. 9 plotted against specimen strain gage data.
has occurred, the stresses cannot be derived assuming a
linearly elastic behavior. Furthermore, the strain would
appear to be accumulated at local rates within the shear
bands of approximately 30,000
y
s. It has been postulated
that strain rates of this order or higher might occur inside
a shear band during dynamic deformation. Although it is
unreasonable to assume that the strain gage could resolve
strains inside a shear band, it does indicate that there is
some type of post-yield behavior such as shear band
deformation occurring.
While the strain derived from the transmitted sig-
nal is affected by dispersion, the strain derived from
the reflected signal does not appear to be as sensitive
(Fig. 14). In fact, up to yielding the two signals appear
to be identical. This would indicate that although the
yield stresses in Figs. 7 and 13 might be affected by
dispersion at strain rates above 3000
y
s, the strain rates
should be reasonably accurate. Given that the typical
reflected signal is similar to the incident signal, as seen
in Fig. 3, with a period twice as large as that of the
transmitted signal, it appears reasonable that the reflected
signal would be insensitive to dispersion.
FIG. 14. Dynamic strain response from the specimen strain gage and
derived from the reflected signal for a strain rate of 3200
y
s.
510
J. Mater. Res., Vol. 11, No. 2, Feb 1996
H. A. Bruck
et al.:
The dynamic compressive behavior of beryllium bearing bulk metallic glasses
VIII. CONCLUSIONS
For the first time, dynamic tests have been conducted
on a metallic glass at strain rates on the order of
10
2
–10
3
y
s. From these tests, it appears that the metallic
glasses retain their elastic-perfectly plastic material be-
havior at high strain rates. Furthermore, the yield stress
of the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
metallic glass was
not found to depend on strain rate. This result is not
unexpected since the micromechanisms for yielding in
metallic glasses are much different than in conventional
polycrystalline metals and ceramics, where the growth
and motion of defects like dislocations and microcracks
are sensitive to strain rate.
Thermal measurements of shear banding processes
in metallic glasses have been made for the first time.
These measurements indicate that temperature increases
due to adiabatic heating occur only after the onset of
inhomogeneous deformation. Furthermore, temperatures
near the melting point of the Zr
41.25
Ti
13.75
Cu
12.5
Ni
10
Be
22.5
metallic glass may be approached within shear bands af-
ter the specimen has failed. This is consistent with SEM
observations of a vein-like morphology on the failure
surface which is attributed to localized melting during
shear banding failure. Since only a single detector was
used for the thermal measurements, it was impossible to
identify the detailed nature of the thermal distribution
near the shear band.
REFERENCES
1. G. I. Taylor, Proc. R. Soc. London
A194,
289 (1948).
2. H. Kolsky, Proc. R. Soc. London
B62,
676 (1949).
3. A. Peker and W. L. Johnson, Appl. Phys. Lett.
63,
2342 (1993).
4. H. A. Bruck, T. Christman, A. J. Rosakis, and W. L. Johnson,
Scripta Metall. Mater.
30,
429 (1994).
5. C. F. Cline and J. E. Reaugh, J. de Phys. IV
1,
139 (1991).
6. A. T. Zehnder and A. J. Rosakis, J. Mech. Phys. Solids
39,
385
(1991).
7. A. J. Rosakis, J. J. Mason, and G. Ravichandran, unpublished.
8. P. S. Follansbee,
Metals Handbook
(American Society of Metals,
Metals Park, OH, 1985), Vol. 8, p. 198.
9. H. J. Leamy. H. S. Chen, and T. T. Wang, Metall. Trans.
3,
699
(1972).
10. C. A. Pampillo and A. C. Reimschuessel, J. Mater. Sci.
9,
718
(1974).
11. G. Ravichandran and S. Ghatuparthi, J. Am. Ceram. Soc.
77,
263
(1994).
12. J. M. Staehler, W. W. Predebon, B. J. Pletka, and J. Lankford,
J. Am. Ceram. Soc.
76,
536 (1993).
13. R. A. Davies, Philos. Trans. R. Soc. London
A240,
375 (1948).
14. P. S. Follansbee and C. Frantz, J. Eng. Mater. Technol.
105,
61
(1983).
15. H. Kolsky,
Stress Waves in Solids
(Dover Publications, New York,
1963).
J. Mater. Res., Vol. 11, No. 2, Feb 1996
511