Published February 1984
| Published
Journal Article
Open
Energy Resolution Enhancement of Mercuric Iodide Detectors
Abstract
A pulse processing technique has been developed which improves the gamma-ray energy resolution of mercuric iodide detectors. The technique employs a fast (100 ns) and a slow (6.4 µs) pulse height analysis to correct for signal variations due to variations in charge trapping. The capabilities of the technique for energy resolution enhancement are discussed as well as the utility of the technique for examining the trapping characteristics of individual detectors. We have achieved an energy resolution of 2.6% FWHM at 662 keV with an acceptance efficiency of 100% from a mercuric iodide detector which gives 8.3% FWHM using standard techniques.
Additional Information
© 1984 IEEE. This work was supported in part by the Jet Propulsion Laboratory Director's Discretionary Fund, NASA grant NGR 05- 002-160, and by U.S. Department of Energy Contract DEAC08- 83NV10282. We thank the High Energy Astrophysics Group at the Jet Propulsion Laboratory under the direction of Alan Jacobson for their contributions to the project. The support and contributions of Profs. Stone and Vogt and W. Althouse at the California Institute of Technology are gratefully acknowledged.Attached Files
Published - 1983-28.pdf
Files
1983-28.pdf
Files
(1.9 MB)
Name | Size | Download all |
---|---|---|
md5:a63a3a77e89b3a3c1420bf4539fdde82
|
1.9 MB | Preview Download |
Additional details
- Eprint ID
- 45268
- Resolver ID
- CaltechAUTHORS:20140429-083600899
- JPL Director's Discretionary Fund
- NASA
- NGR 05-002-160
- Department of Energy (DOE)
- DE-AC08-83NV10282
- Created
-
2014-04-30Created from EPrint's datestamp field
- Updated
-
2021-11-10Created from EPrint's last_modified field
- Caltech groups
- Space Radiation Laboratory
- Other Numbering System Name
- Space Radiation Laboratory
- Other Numbering System Identifier
- 1983-28